Units Selection - Laversab 6300 User Manual

Hide thumbs Also See for 6300:
Table of Contents

Advertisement

Target Value Jog
Each of the three target values can be jogged up or down in small increments by using the
arrow keys. This is done to help the operator achieve a perfect reading on the instrument in
the aircraft, by bumping the set-point in small increments.
To jog a value, select the value by pressing the appropriate target value key and then use the
arrow keys. The selected target value can be "jogged" with the arrow keys until any key
other than the arrow keys is pressed. Each time the arrow key is pressed, the selected target
value will be incremented (UP arrow) or decremented (DOWN arrow) by a specific amount,
depending on the units of the target value.
The increments for the different units are as follows:
Pt units
Knots: 1
mach:
PtinHg: 0.001
QcinHg: 0.001
Ptmb:
Qcmb: 0.01
mph:
kmph: 1
EPR:
Use of the HOLD key
While the Pitot and Static outputs are moving towards their targets, the HOLD key may be
pressed to hold the outputs at their current Actual value. This allows the user to indefinitely
hold the outputs at their current value (regardless of leaks) while attending to other issues.
When the "Hold" is in effect, this is indicated by all the Target values continuously being
blinked (flashing on/off). To release the hold and continue ramping towards the target, the
HOLD key needs to be pressed again.

3.2.3 UNITS SELECTION

Note: Selected Units are maintained through a power OFF/ON cycle
Pitot Units: There are nine units to select from on the Pt output. The units are: knots, Mach,
Pt inHg, Qc inHg, Pt mbar, Qc mbar, miles per hour (mph), kilometers per hour (kmph), and
0.001
0.01
1
0.001
Ps units
Feet: 1
Psin: 0.001
mtrs: 1
mbar: 0.01
22
Climb units
feet/min:
10
inhg/min: 0.01
mtrs/min: 10
mbar/min: 0.1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents