Automatic Zero And O Calibration (Optional); Calibration Due Date (Optional) - BW Technologies GasAlertMicroClip XT User Manual

Gas detector
Hide thumbs Also See for GasAlertMicroClip XT:
Table of Contents

Advertisement

GasAlertMicroClip
User Manual
Automatic Zero and O
Calibration (optional)
2
6. Auto-Zero on Startup: If enabled, the H
sensors are automatically zeroed during startup. Each sensor is
enabled individually. The auto-zero option is enabled for each
sensor upon shipment.
O
Auto-Calibration on Startup: If enabled, the O
2
automatically calibrated during startup. The auto-calibration
option is enabled for the O
If the O
Auto-Calibration on Startup option is enabled, and the
2
Auto-Zero on Startup option is disabled for all sensors, the
following screen displays.
12
S, CO, and LEL
2
sensor is
2
sensor upon shipment.
2
If oxygen is configured to measure 20.8% vol., the oxygen
calibration screen displays 20.8% O

Calibration Due Date (optional)

7. The following screen displays the number of days remaining
before calibration is due. The number of days that displays is
when the next sensor calibration should be performed.
If the Calibration Interval option is defined as 0, the calibration
due date is bypassed during startup.
Note
.
2
Note