Calibration Procedure; Start Calibration; Auto Zero And Oxygen Sensor Calibration - BW Technologies GasAlertMicro 3 User Manual

H2s, co, o2, so2, combustibles gas detectors
Table of Contents

Advertisement

Calibration Procedure

To calibrate the detector and set its alarm setpoints, perform
the following:

Start Calibration

Note: To quit at any point after auto zero, press A. The
detector retains any saved values, and the audible alarm
beeps 4 times before the detector returns to normal
operation.
Note: At any point after auto zero but before calibration gas
is detected, press
to bypass calibration and proceed to
C
alarm setpoints.
Note: Verify that the calibration gas you are using matches
the span concentration value(s) in the detector. See the
section
"Set Span Concentration
1. In a clean atmosphere, press C and H simultaneously
and hold for 5 seconds. The detector beeps 4 times and
the display shows:
Values".
Calibration and Setting Alarm Setpoints
The audible alarm then beeps once signifying that calibration
has begun.

Auto Zero and Oxygen Sensor Calibration

The display flashes N while the detector automatically
zeroes the toxic sensors and calibrates the Oxygen sensor.
Auto Zero
The audible alarm then beeps twice.
Note: Do not apply the calibration gas until the display
shows
K
; otherwise, the auto zero step will fail.
GasAlertMicro
Sensor Fail
29

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gasalertmicro 2Gasalertmicro 1Gasalertmicro 4

Table of Contents