Calibrating With The Ir Link; Calibration Procedure - BW Technologies GasAlertMicroClip XT User Manual

Gas detector
Hide thumbs Also See for GasAlertMicroClip XT:
Table of Contents

Advertisement

Table 12: Connecting to the IR Link
Item
Description
1
Detector
2
IR and charger interface
3
IR Link

Calibrating with the IR Link

a Caution
Only calibrate in a fresh air environment and in a safe area. Do
not calibrate the detector during or immediately after
charging.
To calibrate the detector with the IR Link, complete the following
procedure:
1. Complete steps #1-10 under Setting Span Gas Concentration
Values.
2. Click on the Device Operations tab.
3. Click
. The detector begins calibration. Refer to Auto
Calibrate
Zero and Oxygen Sensor Calibration.

Calibration Procedure

Only calibrate in a fresh air environment and in a safe area. Do
not calibrate the detector during or immediately after
charging.
1. Press and hold
. The detector performs the OFF countdown.
C
Continue holding
as the detector briefly deactivates.
C
a Caution
2. The detector then reactivates and performs the CAL countdown.
Continue holding
until the CAL countdown is complete.
C
If
is not held for the entire countdown, the detector will
C
deactivate.
GasAlertMicroClip
Calibration
Note
39

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gasalertmicroclip xlGasalertmicroclip x3

Table of Contents