19. If the cDAQ chassis is not reserved automatically, select the chassis and click the Reserve
Chassis button. Refer to the
20. Self-test your chassis in MAX by expanding Devices and Interfaces, right-clicking
NI cDAQ-<model number>, and selecting Self-Test. Self-test performs a brief test to
determine successful chassis installation. When the self-test finishes, a message indicates
successful verification or if an error occurred. If an error occurs, refer to
support/daqmx
21. Run a Test Panel in MAX by expanding Devices and Interfaces» NI cDAQ-<model
number>, right-clicking your C Series module, and selecting Test Panels to open a test
panel for the selected module.
If the test panel displays an error message, refer to
Click Close to exit the test panel.
When in use, the cDAQ chassis may become warm to the touch. This is
Note
normal.

Wiring Power to the cDAQ Chassis

Caution
to a DC mains supply or to any supply requiring a connecting cable longer than 3 m
(10 ft). A DC mains supply is a local DC electricity supply network in the
infrastructure of a site or building.
The cDAQ chassis requires an external power source as described in the Power Requirements
section of the specifications document for your cDAQ chassis. Some suggested NI power
supplies are listed in Table 1-8. The cDAQ chassis filters and regulates the supplied power and
provides power to all of the modules. The green POWER LED on the front panel identifies when
the power input is in use.
Complete the following steps to connect a power source to the cDAQ chassis.
1.
Make sure the power source is turned off.
2.
If connected, loosen the connector screw flanges and remove the power screw terminal
connector plug from the cDAQ chassis. Figure 1-9 shows the terminal screws, which secure
the wires in the screw terminals, and the connector screw flanges, which secure the
connector plug on the front panel.
Reserving the Chassis in MAX
.
To ensure the specified EMC performance, do not connect the power input
NI cDAQ-9181/9184/9188/9191 User Manual
section for more information.
ni.com/support