Position 0; Position 1; Position 2; Position 4 - Siemens sitrans lc 300 Instruction Manual

Hide thumbs Also See for sitrans lc 300:
Table of Contents

Advertisement

Position 0

Display

Position 1

Display
Up
Button
Down
Button
Both
Buttons

Position 2

Display
Up
Button
Page 16
LCD displays the PV (Primary Variable) in pF.
In case of a system fault, the display alternates between PV value
and Flt. View the fault details in menu 4.
Pressing either or both push-buttons in menu 0 has no effect.
LCD displays the LRV (Lower Range Value) in pF, occurring when the
range is at 0% and the loop-current is set to 4 mA.
Pressing Up button for less than 1 second adjusts the LRV in the
current step size (initially 0.01 pF).
Holding the Up or Down buttons for more than 1 second increases
the step size to 0.1 pF. If you continue to hold the button, the step size
increases to 10, 100, and 1000 (displayed as 1E3). When no button is
pressed for 4 seconds, the step size decreases to the next smallest
value. At each step size, press the buttons for less than 1 second to
adjust the value.
Pressing Down button for less than 1 second decreases the LRV in
the current step size (initially 0.01 pF).
When held for more than 1 second, the Down Button will increase
the step size by 0.1, 10, 100, and 1000. When no button is pressed for 4
seconds, the step size decreases to the next smallest value.
Pressing both buttons for more than 1 second sets the LRV to the
current PV value.
LCD displays the URV (Upper Range Value) in pF, occurring when the
range is at 100% and the loop-current is set to 20 mA.
Pressing Up button for less than 1 second adjusts the URV in the
current step size (initially 0.01 pF).
Holding the Up or Down buttons for more than 1 second increases
the step size to 0.1 pF. If you continue to hold the button, the step size
increases to 10, 100, and 1000 (displayed as 1E3). When no button is
pressed for 4 seconds, the step size decreases to the next smallest
value. At each step size, press the buttons for less than 1 second to
adjust the value.
SITRANS LC 300 – INSTRUCTION MANUAL
7ML19985HE01

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents