Dräger drager X-dock 5300 Instructions For Use Manual page 22

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 16
Use
The following tests are preconfigured:
Test 1: QT
Fast bump test including alarm testing
Extended bump test incl. zero-point check and
Test 2: EXT
alarm testing.
Calibration, alarm test, fresh air flushing and
Test 3: CAL
certificate.
1. If necessary, open the test gas cylinders.
2. If necessary, switch on the X-dock.
3. Perform a visual inspection of the gas measurement
systems (see chapter 5.1 on page 20).
4. Insert the instruments into the modules (see chapter 5.2 on
page 21).
If Single mode is activated:
The preset test is started automatically by closing the
module cover.
The Status LED flashes blue.
The individual test phases are displayed.
If Favorites mode is activated:
Select required test from favourites bar.
The test will be started automatically.
The Status LED flashes blue.
The individual test phases are displayed.
If Test scheduler mode is activated:
If necessary, log user out on the station (see chapter 4.5 on
page 18).
The preset test is performed according to the configured
time schedule.
If Log-in mode is activated:
Log user in on the station (see chapter 4.5 on page 18).
Select the desired test from favourites bar.
Test will be started automatically.
The status LED flashes blue.
The individual test phases are displayed.
Test passed:
Confirmation is shown on the display.
The Status LED flashes green.
22
0 0
01033286.eps
If required, select the desired instrument field for additional
information.
Remove the instrument from the module.
Test passed, with restrictions:
The status means that partial tests of the favourite could not be
performed because of special settings.
Confirmation is shown on the display.
The Status LED flashes yellow.
If required, select the desired instrument field for additional
information.
Remove the instrument from the module.
Test not passed:
An error message is shown on the display.
The Status LED flashes red.
If required, select the desired instrument field for additional
information.
Identify and rectify the error.
Repeat the test if necessary.
Overview of status LEDs
Colour
Status
blue
flashing
green
flashing
yellow
flashing
red
flashing
0 0
01133286.eps
0 0
01133286.eps
Meaning
Process in progress
Test passed
Test passed, with restrictions
Test failed/cancelled
Dräger X-dock 5300/6300/6600

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

X-dock 6600X-dock 6300

Table of Contents