The following gas measuring devices can be used with the
X-dock and its modules:
X-dock 5300/6300/6600
with Pac module:
Dräger Pac 3500
Dräger Pac 5500
Dräger Pac 7000
2.4
GPL (General Public Licence)
Some of the device software includes open-source software,
which has been published under GPL, LGPL or another open-
source licence. They include GPL GPLv2, LGPL, MIT,
PostgreSQL, Apache, Apache 2, zlib. The source texts of the
software can be obtained from Dräger on a CD under material
number 83 21 874 for at least three years after purchase of the
software. The licence terms and conditions of the software are
included on CD.
3
Installation
NOTICE
i
i
Ensure adequate space for the entire assembly.
The master and all modules must have the same
firmware version. If this is not the case, a firmware
update needs to be carried out (see chapter 6.2 on
page 23).
1. If applicable, install modules on the master station in
accordance with the assembly instructions (X-dock 6300/
6600 only).
A maximum of 10 modules can be installed on one
master station.
The available modules can be combined in any way
desired.
2. If applicable, install wall mount or cylinder holder in
accordance with the assembly instructions.
3. Remove the sleeves from the intended gas inlets and from
the gas outlet.
NOTICE
i
i
If the sleeve is not removed from the gas outlet, the
station will be unable to conduct the self-test without
errors.
4. Fit the gas feed hoses to the gas inlets on the master and
connect them to the regulator valve on the test gas
cylinder.
NOTICE
i
i
Dräger recommends not exceeding a hose length of
10 m for the gas feed hoses.
5. If required, connect an exhaust hose (max. 10 m long) to
the exhaust outlet.
6. Ensure a supply of compressed air or fresh air:
Dräger X-dock 5300/6300/6600
with X-am 125 (+) module:
Dräger X-am 1700
Dräger X-am 2000
Dräger X-am 2500
Dräger X-am 5000
Dräger X-am 5600
Connect the compressed air hose to the compressed
air port (outlet pressure of pressure regulator valve
0.5 bar, flow rate >3 l/min).
OR
If required, connect the fresh air hose to the fresh air
filter.
7. Connect the power pack.
Station with up to 3 modules: Power pack 24 V / 1.33 A
Station with 4 to 10 modules: Power pack 24 V / 6.25 A
The entire system is supplied with power via the master.
NOTICE
i
i
Dräger recommends the use of Dräger gas cylinders
and Dräger pressure regulator valves (see order list).
Alternatively there is the option of using a suitable
pressure regulator valve with 0.5 bar outlet pressure
and >3 l/min flow rate.
Dräger recommends connecting an exhaust hose
(max. 10 m long) to the exhaust outlet to discharge the
test gas into the open air.
4
Basics
4.1
Switching the station on or off
NOTICE
i
i
If no action has occurred for 10 minutes, the station
switches automatically to standby mode.
To switch the station on:
Press and hold the
1 second.
The following information is displayed during the switch-on
process:
Software version number
To switch the station off:
Press and hold the
3 seconds.
The station switches off.
Standby mode:
Standby mode is activated after approx. 10 minutes with no
activity on the station (input via touchscreen or opening/
closing of a module cover).
When the station switches over to standby mode, any user
logged in will be logged off automatically. The user will need
to log back in again when switching back to operating mode.
The touchscreen is switched off in standby mode.
The charging function of X-am 125+ modules with charging
function is not affected by standby mode. The charging
process is not interrupted.
To switch to operating mode:
Tap the function key or
touch the touchscreen or
open or close a module flap.
Installation
key on the master for approx.
key on the master for approx.
17