2. Check the battery contacts and sensor inputs for dirt.
NOTICE
i
i
Devices that have not passed the visual inspection
must not be inserted into the station. Otherwise the
test cannot be fully evaluated correctly.
5.2
Inserting or removing the gas measuring
device into or from a module
To insert the instrument in the module:
1. If necessary, push the lock up slightly and open the module
cover upwards.
2. Place the instrument in the corresponding module.
3. Close the module cover.
The instrument will be detected automatically.
X-am-125+ module with charging function only:
After the instrument is inserted, the charging status is
displayed for approx. 5 seconds via the charging status
LED.
The charging function starts automatically approx. 15
mintues after the last test.
To remove the instrument from the module:
1. Push the lock up slightly and open the module cover
upwards.
2. Remove the instrument.
5.3
X-am 125+ module with charging function
(optional)
WARNING
!
Risk of explosion! Do not charge underground or in
explosion hazard areas. The X-am 125+ modules with
charging function are constructed in compliance with
regulations for fire-damp weather and explosion
protection.
It is only possible to charge the instrument batteries using
the X-am 125+ module with charging function.
The charging time is approx. 4 hours for a completely
empty battery.
Dräger X-dock 5300/6300/6600
A new NiMH supply unit reaches its full capacity after 3 full
charge/discharge cycles. Never store the device for
extended periods (max. 2 months) without a power supply
as the internal buffer battery will run down.
If an error occurs:
Remove from the module and re-insert.
If this does not correct the error, have the module repaired.
!
Overview of charge status LEDs
Colour
2
green
1
green
red
5.4
00233286.eps
A self-test is carried out:
When the station is started up.
When the last successful self-test was more than 24 hours
ago and a test is being carried out.
The station is tested for leaks, pump function, and the software
version of the individual modules and the master.
5.5
!
i
i
CAUTION
A short-circuit in the charging contacts in the modules,
e. g. due to metal objects falling into the device, will not
damage the station but should be avoided due to a
potential risk of overheating and error displays on the
module.
Status
Meaning
on continuously
Charge status 100 %
flashing
Battery charging.
flashing
Charging error

Station self-test

Conducting a test

WARNING
When using methane, propane or butane in the range
>100 %LEL, an exhaust hose (max. 10 m long) must
be connected to the exhaust outlet to ensure the
extraction of excess explosive gas.
NOTICE
Single mode is activated by default.
Several tests can be started and executed in parallel in
the Single mode.
A failure of a LED, horn or vibration test results in a
negative evaluation of the overall test, and thus in the
locking of the respective gas measurement system.
A testing of the sensor reserve is only performed with
sensors that support this function. The results are
displayed under test details and give information about
the state of the sensor.
Use
21