Support Representative to the site or by authorizing the Customer to return the defective accessory or instrument to Varian or to send it to a designated service facility. The Customer shall be responsible for loss or damage in transit and shall prepay shipping cost.
Spare Parts Availability Service Availability It is the policy of Varian to provide operational spare parts for any instrument and Varian provides a variety of major accessory for a period of five (5) years after shipment of the final production services to support its customers run of that instrument.
If this manual is not in your nature language and if you have problems understanding the text, we advise you to contact your Varian office for assistance. Varian cannot accept responsibility for any damage or injury caused by misunderstanding of the information in this manual.
Page 8
Indicates instrument contains parts that can be damaged by Static discharge electrostatic discharge. Take care for proper grounding Warning before handling. Touching this item may result in damage to the instrument Do not touch or personal injury. Page: 2 User manual CP-4900 Micro-Gasifier Varian, Inc.
This instrument is designed for use in combination with the CP-4900 Micro-GC. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Page 10
10. Never try to repair or replace any component that is not described in this manual without the assistance of a Varian service engineer. Unauthorized repairs or modifications will result in rejection of warranty claims.
Gasifier used in combination with the CP-4900 Micro-GC. The Micro-Gasifier is a helping hand to allow sampling of pressurized gases as well as liquefied gases (LPG) to the VARIAN Micro-GC. It is a necessary device for these samples, since the Micro-GC only accepts gaseous samples having a maximum pressure of 100 kPa.
50 ppm to 100 % Sample carry-over < 1 % rsd (as measured for hexane) Varian, Inc is not responsible for printing errors and reserves the right to change specifications, designs and prices without notice and liability. Page: 2 User manual CP-4900 Micro-GC Micro-Gasifier...
For optimal operation, the Micro-Gasifier must be located as close as possible to the sample inlet at the back of the instrument using the heated sample line connection. The CP-4900 Micro-GC must be configured for constant flow operation. More details see CP-4900 Micro-GC user manual “user settings tab”.
Page 14
Introduction and initial operation 1. Remove the two (2) Torx-T20 screws at the back of the CP-4900 Micro-GC. Torx-T20 screws 2. Mount the Micro-Gasifier at the back of the Micro-GC using the two (2) included screws. Taking care of the following notes: - Lead the isolated tubing coming from the Gasifier with some care through the hole at the rear into the heated sample line housing.
Page 15
Micro-GC. The vent control needle valve allows flushing of the regulator under increased flow conditions. The various flows can be measured using a flow meter. 7. Connect the power supply using the included special cable. Varian, Inc. User manual CP-4900 Micro-GC Micro-Gasifier Page: 5...
As a rough guideline, the bomb must be filled with 80% liquid. Position the sample bomb or cylinder to ensure the liquid phase to leave the cylinder or bomb. Page: 6 User manual CP-4900 Micro-GC Micro-Gasifier Varian, Inc.
Page 17
(50 bars) pressures. Alternatively, but a more expensive solution, is the use of a moving piston cylinder. Correct adjustment of the needle valves. Are you sure, that the cylinder or sample bomb is filled with liquid-gas? Varian, Inc. User manual CP-4900 Micro-GC Micro-Gasifier Page: 7...
Introduction and initial operation Cleaning instructions To keep the Power Supply surface clean, refer to the remarks given below: • Switch the CP-4900 Micro-GC off. • Remove the power cable from the mains. • Remove the Micro-Gasifier power cable. •...