Summary of Contents for Varian CP-4900

  • Page 1 4330 EA Middelburg Mulgrave, Victoria 3171 94598 California, USA The Netherlands Australia Tel: ++(1)9259392400 Tel: ++(31)118671000 Tel: ++(61)395607133 Fax: ++(1)9259452360 or Fax: ++(31)118623193 Fax: ++(61)395607950 ++(1)9259452344 © 2008 Varian, Inc. All Rights Reserved. Printed in the Netherlands CP501289 Rev:5 December 2008...
  • Page 2: Safety Information

    If this manual is not in your native language or if you have problems understanding the text, we advise you to contact your Varian office for assistance. Varian cannot accept responsibility for any damage or injury caused by misunderstanding of the information in this manual.
  • Page 3: Table Of Contents

    Sealed radioactive source in the DMD................6 Gas samples.........................6 Carrier gas connection....................6 Operation after Long Storage ..................7 Outputs .........................7 Network Requirements of the CP-4900 Micro-GC ............7 Minimum Computer requirements for all CP-4900 Micro-GC systems ......7 Varian, Inc. Part number: CP501289 Page: 3...
  • Page 4: Pre-Installation Requirements

    Micro-GC equipment. The table below lists the physical dimensions and weight of the CP-4900 Micro-GC and the peripheral instruments which may be installed near it. Allow 10-20 cm of space at the sides and rear of the CP-4900 Micro-GC to permit free air circulation. Height...
  • Page 5: Power Requirements

    In-house tests have shown that the use of an air compressor in combination with a Parker Balston Type 75-45 FTIR purge gas generator provides a constant air quality that allows proper functioning of the DMD. This filter is not supplied via Varian, but should be ordered via the local Parker representative or agent.
  • Page 6: Internal Gas Supply Of The Fieldcase

    (External filter kit part number: CP736729) Option: Genie filter Sample condition: non-condensing gas ambient to 110°C Sample pressure CP-4900 systems: between 0 and 100 kPa Sample pressure Preconcentrator EDU-Varian: Atmospheric pressure Outlet of sample container must fit a stainless steel capillary of 1/16" outside diameter, provided with a Swagelok ®...
  • Page 7: Operation After Long Storage

    The GC can be plugged into any 10BASE-T or 10/100BASE-TX compatible HUB or Switch. • TCP-IP should be used on the Network. Minimum Computer requirements for all CP-4900 Micro-GC systems • Pentium processor > 1.5 GHz. • Ram > 512 MB •...