Varian CP-4900 Micro-GC User Manual

Varian CP-4900 Micro-GC User Manual

Rugged, compact, lab-quality analyzer
Hide thumbs Also See for CP-4900 Micro-GC:
Table of Contents

Advertisement

Quick Links

CP-4900 Micro-GC
North/South America
2700 Mitchell Drive
Walnut Creek
94598 California, USA
Tel: ++(1)9259392400
Fax: ++(1)9259452360 or
++(1)9259452344
© 2002-2009 Varian, Inc. All Rights Reserved
User Manual
Europe
P.O. Box 8033
4330 EA Middelburg
The Netherlands
Tel: ++(31)118671000
Fax: ++(31)118623193
Printed in the Netherlands
Varian BV
Herculesweg 8
4330 EA Middelburg
The Netherlands
Australia/East Asia
679 Springvale Road
Mulgrave, Victoria 3171
Australia
Tel: ++(61)395607133
Fax: ++(61)395607950
CP501265 Rev:7 August 2009

Advertisement

Table of Contents
loading

Summary of Contents for Varian CP-4900 Micro-GC

  • Page 1 4330 EA Middelburg Mulgrave, Victoria 3171 94598 California, USA The Netherlands Australia Tel: ++(1)9259392400 Tel: ++(31)118671000 Tel: ++(61)395607133 Fax: ++(1)9259452360 or Fax: ++(31)118623193 Fax: ++(61)395607950 ++(1)9259452344 © 2002-2009 Varian, Inc. All Rights Reserved Printed in the Netherlands CP501265 Rev:7 August 2009...
  • Page 2: Varian Analytical Instrument Warranty

    Support Representative to the site or by authorizing the Customer to return the defective accessory or instrument to Varian or to send it to a designated service facility. The Customer shall be responsible for loss or damage in transit and shall prepay shipping cost.
  • Page 3: Declaration Of Conformity

    EN 61326-A1 47CFR part 15 ANSI C63.4-1992 Type of Equipment: Micro Gas Chromatograph Model: CP-4900 MANUFACTURER - EU Print Name: G. A. Wassink Company Name: Varian B.V. Herculesweg 8 Address: Signed: P.O. Box 8033 Position: Quality Manager 4330 EA Middelburg Date: November 28, 2001...
  • Page 7: Safety Information

    If this manual is not in your native language and if you have problems understanding the text, we advise you to contact your Varian office for assistance. Varian cannot accept responsibility for any damage or injury caused by misunderstanding of the information in this manual.
  • Page 8 Indicates instrument contains parts that can be damaged by Static discharge electrostatic discharge. Take care for proper grounding Warning before handling. Touching this item may result in damage to the instrument Do not touch or personal injury. Page II User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 9: General Safety Precautions

    It is the responsibility of the Customer to inform Varian Customer Support Representatives if the instrument has been used for the analysis of hazardous biological, radioactive, or toxic samples, prior to any instrument service being performed or when an instrument is being returned to the Service Center for repair.
  • Page 10 13. Never try to repair or replace any component that is not described in this manual without the assistance of a Varian service engineer. Unauthorized repairs or modifications will result in rejection of warranty claims.
  • Page 11: Spare Parts Availability

    SPARE PARTS AVAILABILITY SERVICE AVAILABILITY It is the policy of Varian to provide operational spare parts for any instrument and Varian provides a variety of major accessory for a period of five (5) years after shipment of the final production services to support its customers run of that instrument.
  • Page 13: Table Of Contents

    Space requirements......................7 Power source........................7 Gas supply..........................8 External gas supply ....................8 Safety regards ......................8 ..........................8 AS SAMPLES CP-4900 MICRO-GC INSTALLATION ..................9 ..........................9 NSPECTION ..........................9 NPACKING ..........................10 ACKING LIST CP-4900 Micro-GC ......................10 Accessories ........................
  • Page 14 ETECTOR TCD............................. 36 DMD............................ 36 IN/OUTPUTS..........................37 I/O....................... 38 XTERNAL DIGITAL Ready/not ready signal ....................... 38 (COM) ....................39 OMMUNICATION PORTS I/O......................40 XTERNAL NALOG S............ 41 IFFERENCE ETWEEN ORKSTATION OFTWARE PACKAGE Page 2 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 15 Hubs and switches are… ....................50 Hubs ..........................50 Switches ......................... 51 Routers ........................... 52 CP-4900 M ..................53 ICRO ETWORKS Network Requirements of the CP-4900 Micro-GC ............53 FAQ (F ).................... 54 REQUENT UESTIONS ....................55 LOSSARY OF NETWORK TERMS BNC connector ........................55 Crossover Cable .........................
  • Page 16 Stabilization Time ....................... 94 ........................95 LARM VENTS Remote control external relays................... 96 UserDataStore Address...................... 96 Activate Relay if ........................96 ......................... 98 HANNEL DMD ........................ 101 HANNEL ......................... 102 AMPLER ........................103 RACES Page 4 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 17 ARIAN SERLINK APPLICATIONS CP-4900 External Relays ....................114 CP-4900 instrument status including analog inputs ............115 UserDataStore addresses of Varian userlink applications ..........115 ......................... 117 XTERNAL DEVICES CYCLE WITH STATIC (ELECTRONIC) PRESSURE .............. 119 CYCLE WITH ELECTRONIC PRESSURE CONTROL............120 COLUMN MODULE CONDITIONING ..................
  • Page 18 Congratulations and thank you for purchasing the Varian, Inc. CP-4900 Micro-GC. The CP-4900 Micro-GC is a powerful single or dual or quad channel high speed Micro GC suited for analyses of gaseous samples. This instrument is tailored to meet your needs. The columns are installed and the instrument as a whole has been thoroughly tested at our factory.
  • Page 19: Pre-Installation Requirements

    Micro-GC equipment. The table below lists the physical dimensions and weight of the CP-4900 Micro-GC and the peripheral instruments which may be installed near it. Allow 10-20 cm of space at the sides and rear of the CP-4900 Micro-GC to permit free air circulation. Height...
  • Page 20: Gas Supply

    Introduction and Initial Operation Gas supply Gas quality specifications are listed in the reference document “Pre-Installation requirements CP-4900 Micro-GC CP501289”. External gas supply • Carrier gas pressure: Gas bottle provided wit a proper working two-stage pressure assembly to adjust the carrier gas pressure to 550 ± 10 % (80 ± 10 •...
  • Page 21: Cp-4900 Micro-Gc Installation

    CP-4900 Micro-GC Installation INSPECTION The CP-4900 Micro-GC will arrive packed in one large box and one or more smaller cartons. Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to your local Varian office.
  • Page 22: Packing List

    Check the packing list to see if you have received all that you require. Item Part number External sample filter kit CP736729 Filters 5x for external filter assy. CP736467 CD-ROM user manual CP-4900 Micro-GC CP505532 Cable Dsub9 Male-Female 3 meter CP177138 Locking nut CP420200 x 4 Male Luer CP420100 x 4 Front and back ferrule 1/16”...
  • Page 23: Accessories

    Accessories Item Part number Spare filters (5 pieces) for external filter CP736467 Manual CP-4900 Micro-GC (hard copy) CP501265 Manual CP-4900 Micro-GC power supply (hard CP501267 copy) Heated sample inlet elbow with frit CP740434 Ethernet crossover cable 10 meter CP740293 Ethernet crossover cable, 2 meter...
  • Page 24: Installation

    CP-4900 Micro-GC gas inlet pressure (see Carrier_gas_connection on page 17). Carrier gas The most commonly used carrier gas for the CP-4900 Micro-GC is either He or N . The recommended purity for carrier gas is 99.995% minimum. Carrier_gas_connection on page 17 Connect to Power First, connect the power connector to the CP-4900 Micro-GC, then plug the power cord into an appropriate source of power.
  • Page 25: Sample

    Heated Sample line chapter on page Turn power ON Turn on the CP-4900 Micro-GC using the power switch on the lower front of the instrument. The “READY” LED should light (after two minutes) when ready. (see Front_view on page 14).
  • Page 26: Instrument Overview

    Led blinking: Voltage < 10 Volt SAMPLE 1 and SAMPLE 2 Sample gas inlet connector See here for more details on page 18 POWER ON/OFF SWITCH Switch the CP-4900 Micro-GC ON or OFF Page 14 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 27: Back View

    8 CARRIER 1 CONNECTOR POWER CONNECTOR Carrier gas input connector Power IN connector (male). See here for more details on page 17. See here for more details on page 22 Varian, Inc. User Manual CP-4900 Micro-GC Page 15...
  • Page 28: Inside View

    External Analog I/O signals. See here for more details on page 40 DIGITAL I/O Digital in and output signals, like start_stop, ready_out, start_in and much more. See here for more details on page 38 Page 16 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 29: Carrier Gas Connection

    Carbondioxide Nitrogen Butane Carbon monoxide Your CP-4900 Micro-GC is configured either for carrier gas He (or H ) or it is configured for N (or Ar). Make certain that the carrier gas selection in the controlling Workstation corresponds to the carrier gas connected to your CP-4900.
  • Page 30: Sample Gas

    Introduction and Initial Operation SAMPLE GAS The CP-4900 Micro-GC is an analyzer built for the analysis of gases and vapors only. You are advised to prepare a non-condensing gaseous samples standard sample for routine check up of the instrument. Sample pressure should be between 0-100 kPa (0- 15 psi), the temperature between 0 and 40°C ±...
  • Page 31: Heated Injector Systems

    Introduction and Initial Operation Heated Injector Systems Varian, Inc. User Manual CP-4900 Micro-GC Page 19...
  • Page 32: Heated Sample Line

    The metal surfaces of the sample line heater are very hot and could burn your skin. Open the side panel, the heater will be visible. Sample line heater Remove the insulation. Page 20 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 33 Introduction and Initial Operation The sample line connector will be visible. Connect the sample line. Insulate the sample line coming into the CP-4900, this to prevent damage to the in/out coming cables. Must be insulated Varian, Inc. User Manual CP-4900 Micro-GC Page 21...
  • Page 34: Power

    Introduction and Initial Operation POWER The CP-4900 Micro-GC requires 12.0 VDC, 130W maximum. Refer to the User manual power supply for more information part number CP501267. Only use the power supply that has been supplied with the instrument. SHUT DOWN PROCEDURE...
  • Page 35: Long Storage Recovery Procedure

    Introduction and Initial Operation LONG STORAGE RECOVERY PROCEDURE Follow the recovery procedure below if your CP-4900 Micro-GC has been stored for a long period of time • Remove the 1/8” brass nuts & plastic caps from all the vents/carrier gas connections.
  • Page 36: Initial Operation

    90 of the data-handling package manual. Send the instrument parameters from the data-handling package software to the Micro- GC. You can watch the CP-4900 Micro-GC status in the instrument status tab on page INJECTING THE SAMPLE Connect the sample by means of a 1/16" line to the CP-4900 Micro-GC.
  • Page 37: Column Cp-Sil 5 Cb 6 Meter Unheated

    CP-Sil CB columns. Nitrogen, methane, carbon dioxide, and ethane are not separated on these columns. They produce a composite peak. For separation of these components a HayeSep A column is advised. Varian, Inc. User Manual CP-4900 Micro-GC Page 25...
  • Page 38: Column Cp-Sil 5 Cb 4 Meter Heated

    : 150kPa (21PSI) Sample time : 30s Injection time : 40ms Run time : 30s Detector sensitivity : Auto Peak identification Component Concentration Composite Balance Ethane 8.1% Propane 1.0% i-Butane 0.14% n-Butane 0.20% Page 26 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 39: Column Cp-Sil 5 Cb 8 Meter Heated

    : 150kPa (21PSI) Sample time : 30s Injection time : 40ms Run time : 60s Detector sensitivity : Auto Peak identification Component Concentration Composite Balance Ethane 8.1% Propane 1.0% i-Butane 0.14% n-Butane 0.20% Varian, Inc. User Manual CP-4900 Micro-GC Page 27...
  • Page 40: Column Cp Sil13Cb Tbm 12 Meter Heated

    : 40°C Component Concentration Injector temperature : 50°C Methane Balance Column pressure : 250kPa (38PSI) 6.5ppm Sample time : 30s Injection time : 255ms Run time : 80s Detector sensitivity : Auto Page 28 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 41: Column Hayesep 40Cm Heated

    The HaySep A column separates oxygen, methane, carbon dioxide, ethane, acetylene, ethylene, and selected sulfur gases. Nitrogen coelutes with oxygen. Components with a higher molecular weight than propane have long retention times on this column. Maximum allowable column temperature: 160 °C Varian, Inc. User Manual CP-4900 Micro-GC Page 29...
  • Page 42: Column Molsieve 5Å 20 Meter Unheated

    : Auto The Molsieve 5Å column is designed to separate: hydrogen, carbon monoxide, methane, nitrogen, oxygen, and some noble gases. Higher molecular weight components have much higher retention times on this column. Page 30 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 43: Conditioning Of Molsieve Columns

    Micro-GC's frequency of use is very high, you might adopt a standard reconditioning procedure of leaving the instrument with the oven temperature at 180°C overnight. The longer the re+onditioning period the better the column performance without damage. Varian, Inc. User Manual CP-4900 Micro-GC Page 31...
  • Page 44: Column Ppq 10 Meter Heated

    : 110°C Composite Balance Column pressure : 150kPa (21PSI) Ethane 8.1% Sample time : 30s Propane 1.0% Injection time : 40ms i-Butane 0.14% Run time : 50s n-Butane 0.20% Detector sensitivity : Auto Page 32 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 45: Reference

    Reference REFERENCE INSTRUMENT DESCRIPTION CARRIER GAS The Varian, Inc. CP-4900 Micro-GC is configured for the use with either He (or H ) or N (or Ar). Having chosen one option definitely rules out the use of the other without the instrument undergoing internal changes prior to any switching of carrier gas type.
  • Page 46: Micro Electronic Gas Control (Egc)

    VENT MICRO ELECTRONIC GAS CONTROL (EGC) The CP-4900 Micro-GC has built-in Micro EGC regulators that can be adjusted to get a constant or programmed pressure control, which result in a constant or programmed flow through the injector, column and detector. The pressure range of the Micro EGC is between 50-350 kPa (7 - 49 Psi).
  • Page 47: Column

    Table I shows several standard columns as supplied in the CP-4900 Micro-GC and selected applications. Other columns are available by contacting Varian, Inc.
  • Page 48: Detector

    Reference DETECTOR The CP-4900 Micro-GC is equipped with a thermal conductivity detector (TCD). This detector responds to the difference in thermal conductivity between the carrier gas and the sample components. The construction of a TCD is such that the changing thermal conductivity of the carrier gas stream, due to components present, is compared to the thermal conductivity of a constant reference gas stream.
  • Page 49: In/Outputs

    The CP-4900 Micro-GC has several in/output ports accessible inside the instrument for interfacing with external devices. Open the cover. At the front of the CP-4900 Micro-GC, the external device connectors will be visible. Close the cover after connecting the cables! Varian, Inc.
  • Page 50: External Digital I/O

    Reference EXTERNAL DIGITAL I/O Connection between CP-4900 Micro-GC and external devices is made with the appropriate cable to the External Digital I/O port (see also External_devices on page 117). CP-4900 Digital IN1 (not definied) Digital IN2 (not definied) Digital OUT1 (not definied)
  • Page 51: Communication Ports (Com)

    RS-232 RS-232 COM 3 RXA(+) RXB(-) TXB(-) TXA(+) RS-485 Difference between Workstation Software packages, click here on page Varian, Inc. User Manual CP-4900 Micro-GC Page 39...
  • Page 52: External Analog I/O

    +15 VDC out (electronic fused max 100 mA) -15 VDC out (electronic fused max 100 mA) +5 VDC out (electronic fused max 500 mA) +12 VDC out (electronic fused max 500 mA) Analog GND External Analog I/O Page 40 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 53: Difference Between Workstation Software Packages

    External Ready: Yes External Ready: Yes Timed Relay: Yes Timed Relay: Yes Relay Control Alarm Relay: Yes Alarm Relay: Yes Extension Box: Yes Extension Box: No Only one Vici valve possible in Galaxie Varian, Inc. User Manual CP-4900 Micro-GC Page 41...
  • Page 54: The Differential Mobility Detector

    Ni Radiation Safety Manual. General functionality Varian has extended its CP-4900 Micro Gas Chromatography product line with a new detector. The Differential Mobility Detector, DMD, integrated into the CP-4900 chassis is a joint development of Sionex Corporation and Varian Instruments.
  • Page 55 As illustrated in Figure 1, the net voltage applied results from three independent settings: 1. A fixed voltage, or RF-Voltage, 2. A scanning voltage, also referred to as Compensation Voltage, A RF modulated voltage. Varian, Inc. User Manual CP-4900 Micro-GC Page 43...
  • Page 56: Cp-4900 Micro-Gc Dmd Channel

    The application specific method is saved inside the DMD hardware. CP-4900 Micro-GC DMD Channel The DMD channel is always located in the last position (against the backpanel)
  • Page 57: General Operation

    In case of a suspected misalignment of the method, where retention time or compensation voltage may be out of tune, a tuning run may be needed. Please contact for more information your local Varian Inc. subsidiary or Varian, Inc. representative.
  • Page 58: Networking Basics

    Used for larger office networks 10+ Computers. Different types of networks Star Topology • Nodes are connected by branches. • Central Device is usually a hub or Switch. • Central Device controls the communication of each node. Page 46 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 59: Bus Topology

    The chosen cable type depends on the required distance, speed, interference and security. Coaxial • Single Copper connector enclosed by protective braided copper shield. • Widest implementation is broadband access to cable TV. • Must have a terminator at each end. Varian, Inc. User Manual CP-4900 Micro-GC Page 47...
  • Page 60: Coaxial Rg-58

    Cable type Application Analog voice-telephone Digital voice- 1 Mbps data UTP/STP 16 Mbps data UTP/STP 20 Mbps data UTP/STP 100 Mbps data UTP/STP 155 Mbps data (proprietary) UTP/STP 1000 Mbps data (proprietary) Page 48 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 61: Protocols

    The Internet protocol suite Starting IP Ending IP Subnet Mask Type 0.0.0.0 255.255.255.255 Public 10.0.0.0 10.255.255.255 255.0.0.0 Private 172.16.0.0 172.31.255.255 255.255.0.0 Private 192.168.0.0 192.168.255.255 255.255.0.0 Private Note: never use 0.0.0.0. or 255.255.255.255 Varian, Inc. User Manual CP-4900 Micro-GC Page 49...
  • Page 62: Networking Devices (Hubs, Switches, And Routers )

    Pass on traffic to the network regardless of the intended destination. • Result in a lot of redundant and unnecessary traffic sends across a network. • Ideal for smaller networks (less than 30 devices). Page 50 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 63: Switches

    • Reduces the amount of unnecessary traffic. • Is in fact a “Smart” device, which is able to learn. • Is a necessity when networks get larger (more than 50 users). Varian, Inc. User Manual CP-4900 Micro-GC Page 51...
  • Page 64: Routers

    Provide a link between networks with a different logical connection (protocols). • Consist of at least two network interface cards. • Don’t have to be a black box, but could also be a regular PC with routing software. Page 52 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 65: Cp-4900 Micro-Gc And Networks

    Reference CP-4900 MICRO-GC AND NETWORKS Network Requirements of the CP-4900 Micro-GC • Customer Cable type should be Cat3 or Cat5 UTP / STP. • Customer network should comply with Standard Ethernet (IEEE 802.3). • The network must have either 10BASE-T or 10/100BASE-TX compatible HUB or Switch.
  • Page 66: Faq (Frequent Ask Questions)

    Q: Can I control my Micro-GC from anywhere in the world via the Internet? A: Yes, if the customer network is designed for this, and has Internet access or remote access facilities (the ports 4900,4901 and 4902 must be open). Page 54 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 67: Glossary Of Network Terms

    (between 1 and 255) that provide routing information used by the TCP/IP protocol to establish a reliable connection. Without the IP Address, communications would be bogged down trying to establish connections to Ethernet addresses at unknown locations. Varian, Inc. User Manual CP-4900 Micro-GC Page 55...
  • Page 68: Patch Cable

    RJ45-style connectors are used by the CP-4900 Micro-GC. TCP/IP An international standard protocol used by the Internet. We use this protocol for communication to the CP-4900 Micro-GC. You may find several network protocols, such as IPX/SPX and NetBEUI, installed on you computer. Page 56 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 69: Workstation Software

    In that case the instrument will accept the first incoming connection type (serial or Ethernet) and will lock the other connection type. There can never be a connection on the two different ports at the same time. Varian, Inc. User Manual CP-4900 Micro-GC Page 57...
  • Page 70: Cp-Maitre Elite

    Multiple CP-4900s must be controlled serial or Ethernet, not a mixture of these two. The maximum number of CP-4900s controlled by serial communication is two. For Ethernet this is limited by your license. The CP-4900 driver scope Page 58 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 71: Serial Communication

    Reference SERIAL COMMUNICATION To control the CP-4900 Micro-GC via serial communication, just connect the serial cable (supplied with the instrument) to a free serial port of your computer configured as RS232, and plug the other side into port COM1 of the CP-4900 (see...
  • Page 72: Ethernet Communication

    Reference ETHERNET COMMUNICATION An Ethernet crossover cable (RJ-45 connector) is included with the Ethernet option. The crossover cable must be used to control the CP-4900 Micro-GC in a peer-to-peer communication (see Inside View on page 16). This requires IP addresses in the same range for computer and Micro-GC. Only the most right IP segment must be different.
  • Page 73: Multiple Instrument (Local Network)

    CP-Maitre Elite maximum connections are limited by CP-4900 Micro-GC the computer speed, license and network Multiple CP-4900s connected to a local performance. workstation. IP address in same range as computer IP address. Varian, Inc. User Manual CP-4900 Micro-GC Page 61...
  • Page 74: Global Network: Multiple Instruments

    IP address in range of subnet 1 Subnet 2 CP-Maitre Elite IP address in range of subnet 2 CP-4900 in subnet 1 can also be IP address in range of subnet 2 controlled Page 62 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 75: Instrument Configuration

    Configure… from the popup menu. You must have system administration privileges in order to configure instruments. When you select the Configure command for a Varian CP-4900, a configuration dialog box will appear where you start your CP-4900 Micro-GC configuration.
  • Page 76 (such as System Administration Wizards). Select the Varian CP-4900 as Instrument Type from the drop-down list. Then click the Configure button to continue with the Micro-GC configuration. A dialog will appear where you can complete the configuration of the Micro-GC.
  • Page 77: Modules Configuration

    Reference MODULES CONFIGURATION A Varian Micro-GC icon is displayed in the Available Modules box on the left. To add the module for configuration, click the icon in the left box, and then click the Green button to move it to the Configured Modules box on the right, or double-click on the icon. The Analysis Options button is used to initiate any optional analysis software for this instrument.
  • Page 78: Micro-Gc Configuration Setup

    Upload Instrument Configuration on page If there is no CP-4900 Micro-GC connected do not press the “Upload” button but manually select the hardware settings which match a virtual connected CP-4900 Micro-GC. This can be useful for method development on a computer without having a CP-4900 Micro-GC connected.
  • Page 79: Communication Setup

    Reference COMMUNICATION SETUP In the CP-4900 Micro-GC window select the type of communication for instrument control. Ethernet can only be selected if the instrument is equipped with an Ethernet card. Serial communication For serial communication select a free serial port on your computer which is configured as RS232.
  • Page 80: Ehternet Connection Setup

    IP address and exit the “Setup Ethernet Connection” window. It is assumed that the CP-4900 Micro-GC has the right IP address if it is not part of this subnet, otherwise assigned first a new IP address, see topic Assign new IP address.
  • Page 81: Find Instruments On The Network

    Reference FIND INSTRUMENTS ON THE NETWORK The “Find Instruments on the net” is used to view all the CP-4900 Micro-GC on the local subnet. For each CP-4900 Micro-GC detected the IP address, instrument serial number and status is displayed. If the instrument is already controlled from another computer, the IP address of that computer will be displayed.
  • Page 82: Assign New Ip Address Via Ethernet Connection

    For a global network, the system administrator must be contacted. 3. Enter the IP address that must be assigned to the CP-4900 Micro-GC. This IP address must be in the same range as the computers IP address. The three left IP segments must be identical.
  • Page 83: Assign New Ip Address Via Serial Connection

    3. Click on the button “Serial Setup IP-address”, the “Serial Setup Ethernet Settings” window appears. 4. Enter the IP address that must be assigned to the CP-4900 Micro-GC. This IP address must be in the same range as the computers IP address. The three left IP segments must be identical.
  • Page 84: Upload Instrument Configuration

    The instrument hardware settings, user settings, the instrument serial number and general information like software versions numbers, etc. will be uploaded from the CP-4900 Micro-GC and viewed in the Configuration tabs "Hardware” below, “User...
  • Page 85 Reference In the configuring page [Detector] will automatically select “DMD”, upon uploading the configuration for a CP-4900 DMD instrument. At the same time the checkbox for heated injector on Channel 2 disappears. Varian, Inc. User Manual CP-4900 Micro-GC Page 73...
  • Page 86: User Settings Tab

    “None”, 1, 2 and 3 flush cycles. The If it is selected all the GC channels will Flush cycle is invoked at startup of generate a default chromatogram. the instrument or when pressure is restored after a low-pressure error. Page 74 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 87: Cp-Maitre Elite Tab

    Select between kPa and PSI. The column and ambient pressures status will be displayed in the selected pressure units in the Method Setup on page Instrument Control on page 85 window of instrument control. Varian, Inc. User Manual CP-4900 Micro-GC Page 75...
  • Page 88: Automation Tab

    This setting can be used to acquiring auxiliary traces (see Select this setting to activate the synchronize another device with Acquire_auxiliary_traces on page connected Extension Box. the CP-4900 Micro-GC (see 77). page Wait_starting_CP4900_external_ Ready_IN on page 79). Page 76 User Manual CP-4900 Micro-GC...
  • Page 89 Select this setting to store the instrument status into the UserDataStore library. In instrument control the status is continuously stored into the UserDataStore library. The instrument status will become accessible for other applications like the Varian userlink applications PCI1000Com, Mlink32, I/O-Control or any other self-written program. Also the instrument status as stored into the UserDataStore database can be used for external relay setup, see item “Remote control external relays”.
  • Page 90 Vial is set to one. Vial number two result in Relay 2 activated and Relay 1 deactivated. Make sure to select “External Trigger” in the Trigger Tab on page 92 of Instrument Method Setup in case an auto sampler is configured. Page 78 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 91 This setting can be used to synchronize another device with the CP-4900 Micro GC. If selected the CP-4900 driver will postpone the start of the CP-4900 till the Ready-In signal is true. Varian, Inc. User Manual CP-4900 Micro-GC Page 79...
  • Page 92: Vici Valve

    In order to control a VICI electric actuated multi positional valve, the valve identity for this valve must be known. By default the VICI valve, as shipped by Varian, has not an identity. In case only one VICI valve is controlled by the CP-4900 driver, a valve identity is not required, you can skip this chapter.
  • Page 93: Info Tab

    MPU of the CP-4900. in the CP-4900 on every GC analytical module part of the channel. GC channel. I/O Controller Software version of the I/O Controller, a micro controller in the CP-4900. Varian, Inc. User Manual CP-4900 Micro-GC Page 81...
  • Page 94: Remote Firmware Update

    66 window. The CP-4900 Micro-GC as well as the CP-Maitre Elite CP-4900 driver supports remote update of the firmware in the CP-4900. The firmware consists of a number of files. Only the main GC application (GC.EXE) can be updated.
  • Page 95: Pressure Sensor Calibration

    Instrument Status on page 85 window. It’s important to follow exactly the instructions given by the CP-4900 Micro-GC. Prior to the calibration, disconnect the carrier gas tubing. Allow the CP-4900 to stabilize (2 minutes). The TCD filaments will automatically switch off when low carrier gas pressure has been detected.
  • Page 96: Upload Method

    98 “Common tab” on page DOWNLOAD METHOD Once you have completed the Instrument Setup for your CP-4900 Micro-GC, you can transfer the method using the Control/Download Method command. Only the settings from the ”Channel tabs” on page 98 “Common...
  • Page 97: Instrument Status

    INSTRUMENT STATUS The Control/Instrument Status command is used to bring up a real-time status screen for your Varian CP-4900 Micro-GC. The instrument method settings (“Set” column) will only appear if a method is up or downloaded from/to the CP-4900 Micro-GC.
  • Page 98: Extension Box Status (Optional)

    Reference Extension Box status (optional) This shows the states of the eight-extended relays of the Extension box. Page 86 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 99: Show Autosampler Tray

    The Volume value represents the pre-run time or flush time. This is the time between start request and the actual start of a run. If no Auto Sampler configured the “Vial” and “Volume” parameters must be ignored. Varian, Inc. User Manual CP-4900 Micro-GC Page 87...
  • Page 100: Event Configuration

    Method Development section. The external events tab in Instrument Setup will not be available unless you have configured your instrument with one or more external events. Page 88 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 101: Start Sequence

    A preview run does not start an injection or run on the instrument, it only acquires detector data while the auto zero function is switched off. A Preview Run is not time limited. The user has to stop it manually by pressing the Stop icon. Varian, Inc. User Manual CP-4900 Micro-GC Page 89...
  • Page 102: Micro-Gc Method Setup

    Reference MICRO-GC METHOD SETUP Before you can make a run with your CP-4900 Micro-GC, you must set up the instrument and data acquisition parameters for one or both channels of the instrument. To access the Instrument Setup dialog, click the Instrument Setup icon on the toolbar, or select the Method/Instrument Setup command from the menu.
  • Page 103: Aux Traces Main Tab

    The displayed units for analog inputs are only correct if in the Aux Traces Tab on page 103 scale factor is set to one and the offset to zero. It is not possible to change these units. Varian, Inc. User Manual CP-4900 Micro-GC Page 91...
  • Page 104: Trigger Tab

    4900 configuration and the CP-4900 is started by CP-Maitre Elite when the ‘OK’ button is pressed in the appeared dialog. This dialog will appear if a single run on page 87 sequence is started on page 89. CP-Maitre Elite waits till the operator presses ‘OK’. Page 92 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 105: Common Tab

    This is the current continuous flow setting in the CP-4900. At startup of Instrument control this CP-4900. At startup of Instrument control this setting is uploaded from the CP-4900. setting is uploaded from the CP-4900. Varian, Inc. User Manual CP-4900 Micro-GC Page 93...
  • Page 106: Stabilization Time

    If during the stabilizing period one of the channel temperatures or pressures become Not Ready the overall instrument state will jump to Not Ready and the whole process will start again. Enter zero for Stabilizing time if the stabilizing period is not required. Page 94 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 107: Alarm Events Tab

    UserDataStore address. Activate Relay if Here can be chosen between “Within limits”, “Outside Relay Time Event Table limits” and “Equals lower limit” (see Activate_relay_if Click here for more details on page page 96). Varian, Inc. User Manual CP-4900 Micro-GC Page 95...
  • Page 108: Remote Control External Relays

    The relay setup conditions will become operational after the method has been downloaded to the CP-4900 Micro-GC. Every two seconds the CP-4900 driver checks the relay switch condition. If the condition changes the relay will be immediately switched.
  • Page 109 This program requires two program lines. Attention the end of the run will now be postponed with 30 seconds! Event 1=Start of Injection, Time=20, Action=Activate relay 1 Event 2=Post Run, Time=30, Action=Deactivate relay 1 Event 3=None Event 4=None Varian, Inc. User Manual CP-4900 Micro-GC Page 97...
  • Page 110: Channel Tabs

    Low Medium, High or Extra high. acquired detector data. Auto is highly advised as it gives the widest linear dynamic range, with the lowest noise level. Will be continued on the next page Page 98 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 111 If you have entered an incorrect parameter for any of the above items, an error message will be displayed in this field to aid you in correcting the setting. Will be continued on the next page Varian, Inc. User Manual CP-4900 Micro-GC Page 99...
  • Page 112 Here the channel description as Micro-GC is displayed. At startup of Instrument defined in the User Settings control this setting is uploaded from the CP-4900 page 74 of the instrument Micro-GC. configuration is displayed. Page 100 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 113: Channel Tab Dmd

    NOTE: Tuning of the DMD channel is reserved for trained service engineers. The Channel Description in the lower right hand corner indicates the presence of a Sionex DMD Channel (unless the channel description is overwritten in the configuration). Varian, Inc. User Manual CP-4900 Micro-GC Page 101...
  • Page 114: Sampler Tab

    Click on this button to reset the entire auto sampler table to default settings. Transmit after injection checkbox Select this box to enable sending “Transmit after injection” commands to the VICI sampler. Page 102 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 115: Aux Traces Tab

    Remote Access Instrument on page 110 data and Automation tab on page 76 of the instrument configuration. Varian, Inc. User Manual CP-4900 Micro-GC Page 103...
  • Page 116: Eliteexport.dll User File

    Data stored in the UserDataStore database can be used by the CP-4900 driver to control the external relays of the CP-4900 Micro-GC. For additional parameters the ‘/i1’ argument has to be entered. This is the UserDataStore instrument number. Make sure to enter the same instrument...
  • Page 117 Enter here the address to store the defined peak parameter. Recommended is to use addresses in the range 520--1199. TypeParam Enter here the type of peak parameter you want to store for the defined component above. Varian, Inc. User Manual CP-4900 Micro-GC Page 105...
  • Page 118: Standard Control

    EXTERNAL STARTED External start mode is a CP-4900 Micro-GC controlled by CP-Maitre Elite but started by an external device. The following steps are required: 1. Configure the CP-4900. This consist of setup the...
  • Page 119: Auto Sampler Control

    The CP-4900 driver will now check every two seconds the relay switch conditions. If a condition changes the relay(s) will be immediately activated or deactivated depending on the condition. Varian, Inc. User Manual CP-4900 Micro-GC Page 107...
  • Page 120: External Pump Control

    The topic Remote Access Instrument Status on page 110 describes how to transfer the converted signals to the UserDataStore library. This enables other applications to retrieve the instrument status including the converted signals. Page 108 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 121: External Device Synchronization

    1. CP-Maitre Elite is initiating a new run from a sequence. 2. The Method for that particular run is downloaded to the CP-4900 Micro-GC. The instrument will control now to new setpoints. 3. Because the Auto Sampler is configured, the Pre-run phase will be initiated.
  • Page 122: Remote Access Instrument Status

    Reference REMOTE ACCESS INSTRUMENT STATUS The Instrument status of the CP-4900 Micro-GC can be accessible for other applications like Varian userlink applications: Mlink32, I/O Control, PCI1000Com or any other self written program capable to collect instrument status from the UserDataStore database.
  • Page 123 Injector temperature channel 4 in Celsius 1251 Column pressure channel 1 in kPa 1252 Column pressure channel 2 in kPa 1253 Column pressure channel 3 in kPa 1254 Column pressure channel 4 in kPa Varian, Inc. User Manual CP-4900 Micro-GC Page 111...
  • Page 124: General Error State

    Initialization error Internal communication error Instrument EDS incorrect EDS incorrect Internal power failure Flush cycle aborted GC module changed TCD Gain calibrated TCD Offset calibrated Null String ADC reading out of range Page 112 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 125 DMD sensor voltage error DMD RF voltage error DMD Vc voltage error DMD sensor pressure error DMD tuning process error DMD gas flow error DMD system voltage error DMD communication error DMD sensor error Varian, Inc. User Manual CP-4900 Micro-GC Page 113...
  • Page 126: Varian Userlink Applications

    Everything stored in the shared UserDataStore database can be used to setup a switch condition, since every Varian userlink application is capable to store its specific results into this database. To store CP-Maitre Elite 3.2...
  • Page 127: Cp-4900 Instrument Status Including Analog Inputs

    UserDataStore addresses of Varian userlink applications Find below a table of relevant UserDataStore addresses, which can be used, for alarming purposes. The addresses are all part of the Varian Userlink applications: ISOCAL32, BTU32, InstrumentNorm32 and Mlink32. The addresses are specific for each instrument (maximal four instruments)
  • Page 128 Stream Position from CP-Maitre Elite Total ESTD summed of all channels EliteExport.dll or Mlink32 520 and User defined parameters as defined in the UserDataStore sheets output section of Mlink32 or as defined in the EliteExport.ini file. Page 116 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 129: External Devices

    38). 3. External Start In The CP-4900 driver supports the start of the CP-4900 Micro-GC by an external device (Flow computer or PLC). The CP-4900 Micro-GC processes this event (start run) and passes the event to the CP-4900 driver (see also...
  • Page 130 Reference Page 118 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 131: Cycle With Static (Electronic) Pressure

    The Workstationsoftware will time allows the acquisition is run is finished. be downloaded wait until data is received sample to settle in started. via the from the CP-4900 Micro-GC. the sample-loop Workstation Varian, Inc. User Manual CP-4900 Micro-GC Page 119...
  • Page 132: Cycle With Electronic Pressure Control

    The pressure from the final pressure to the initial same. remains the same. pressure. Varian, Inc. User Manual CP-4900 Micro-GC Page 120...
  • Page 133: Column Module Conditioning

    Follow this procedure to make sure that any “water” that might be present inside the analytical column is removed before the TCD is switched on. Also follow this procedure if the CP-4900 Micro-GC or CP-4900 Micro-GC module has been stored for a long period.
  • Page 134: Backflush Option

    The Varian CP-4900 Micro-GC is optionally available with GC modules that incorporate backflush capabilities. A backflush system always consists of a pre-column and an analytical column.
  • Page 135: Description

    (back) flushed to the vent (see backflush diagram). On the analytical column the separation goes on, because as there the flow is not inverted! Varian, Inc. User Manual CP-4900 Micro-GC Page 123...
  • Page 136: Tuning

    Be aware that a small pre-column is used. It is not always possible to cut between 2 peaks! Backflush Time range 0.5 seconds until maximum run time. Special value BACKFLUSH TIME = 0 this puts the system in FOREFLUSH mode during run the entire run. Page 124 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 137: Shipping Instructions

    Reference SHIPPING INSTRUCTIONS If your CP-4900 Micro-GC for any reason must be sent back to the factory, it is very important to follow these additional shipping preparation instructions: 1. Add all the vent caps at the back of the CP-4900 Micro-GC.
  • Page 138: Error List

    During operation a series of events and error messages are generated indicating start or finishing of certain actions and procedures as well as smaller and fatal errors somewhere in the instrument. This section describes how CP-4900 micro-GC reacts to these events or messages.
  • Page 139: Error List

    Instrument EDS incorrect Instrument Electronic Datasheet incorrect Call service EDS incorrect Electronic Datasheet incorrect Call service Internal power failure During/after initialization, internal supplies Call service Flush cycle aborted Flush cycle stopped before completion Varian, Inc. User Manual CP-4900 Micro-GC Page 127...
  • Page 140 Column cleaning Instrument in column cleaning state Equilibrating temperature zones Instrument stabilizing after column cleaning Wait till Ready DMD control to temperature Notification that DMD is warming up. Wait for setpoint completion Page 128 User Manual CP-4900 Micro-GC Varian, Inc.
  • Page 141 Internal Software Error, IOC fatal error 1 Auto reboot Watchdog Error: IOC Fatal error Internal Software Error, IOC fatal error 2 Auto reboot Watchdog Error: IOC Fatal error Internal Software Error, IOC fatal error 3 Auto reboot Varian, Inc. User Manual CP-4900 Micro-GC Page 129...