Error which has
occurred
Switch the device off if an error symbol appears. Check the possible causes
and note the 10 golden rules in chapter 3 and the instructions in chapter
8. Relax for a minute and take the measurement again. Do not move or
talk during the measurement.
7. Significance of symbols on
the device and cuff

7.1 Control displays

Flashes when the device is
measuring and the pulse is
being taken.
Replace batteries if the
battery symbol is perma-
nently illuminated.
Measuring error, cf.
chap. 6.
Possible causes
Moving, talking or
excitement during the
measuring procedure.
Lack of relaxation before
taking a measurement.
Stimulants taken before
measurement.
Remedy
Please take measurements
in a relaxed position whilst
seated. Do not talk or
move during the measur-
ing procedure.
Relax for 5 minutes before
taking a measurement.
Please avoid alcohol /
nicotine and caffeine for
one hour before taking a
measurement.
Appears during inflation.
Appears during automatic
checking.
Displays the stored
M1
measured values for
person 1.
Displays the stored
M2
measured values for
person 2.