Chapter 3 Features and Functions
Amplitude Modulation (AM)
Carrier Waveform Shape
• AM carrier shape: Sine , Square, Triangle, Ramp, or Arbitrary
waveform. The default is Sine.
• You cannot use the noise function or dc volts as the carrier waveform.
• Front-Panel Operation: Press any of the front-panel function keys
except
. For arbitrary waveforms, the
Noise
waveform that is currently downloaded.
• Remote Interface Operation:
FUNCtion:SHAPe {SINusoid|SQUare|TRIangle|RAMP|USER}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset with a single command.

Carrier Frequency

• Carrier frequency: 100
The default is 1 kHz.
• For arbitrary waveforms, the maximum carrier frequency depends on
the number of points specified in the waveform. The five built-in
arbitrary waveforms can be output at a maximum of 5 MHz.
Number of Arb Points
8 to 8,192 (8k)
8,193 to 12,287 (12k)
12,288 to 16,000
• Front-Panel Operation: Press the
frequency for the selected function (the present amplitude and offset
voltage are used).
• Remote Interface Operation:
FREQuency {<
frequency
72
Arb
µ
Hz to 15 MHz (100 kHz for triangle and ramp).
Minimum Frequency
Maximum Frequency
µ
100
Hz
µ
100
Hz
µ
100
Hz
key to select any valid
Freq
>|MINimum|MAXimum}
key selects the
5 MHz
2.5 MHz
200 kHz