Chapter 3 Features and Functions
Calibration Overview
To Unsecure for Calibration You can unsecure the function
generator for calibration either from the front panel or over the remote
interface. The function generator is secured when shipped from the
factory and the security code is set to "HP033120".
• Front-Panel Operation:
1: SECURED
(CAL MENU)
If the function generator is secured, you will see the above command
when you go into the
CAL MENU
level in the menu, you will notice that the "
hidden when the function generator is secured.) To unsecure the
function generator, go to the "
command, enter the security code, and then press Enter.
∧000000:CODE
When you go to the "
COMMANDS
you will notice that the function generator is unsecured. Notice also
that the "
CALIBRATE
perform a calibration.
1: UNSECURED
• Remote Interface Operation:
CALibration:SECure:STATe {OFF|ON},<
To unsecure the function generator, send the above command with
the same code used to secure calibration. For example,
"CAL:SEC:STAT OFF,HP033120"
. (If you move across the "
CALIBRATE
" level of the
PARAMETER
" level in the
" command is no longer hidden and you can
"
COMMANDS
" command is
SECURED
again,
CAL MENU
>
code
119