SITRANS LVS200 Pipe Extended Version

Assembly
Suggested tools:
medium Phillips or 6 to 8 mm (¼") flat screwdriver •
3 mm (1/8") flat screwdriver
wire cutters
wire strippers
enclosure
1.
Open the enclosure lid; remove electronics module.
2.
Lead the sensor cable through the customer supplied 1" tube and enclosure.
3.
Assemble the fork assembly, the pipe extension, and the enclosure using the 36 mm
open end wrench. Seal the pipe threads with an appropriate sealant.
Do not turn fork assembly. Do not bend the fork during assembly.
Note:
4.
Line up the fork and the tine orientation marking as shown in dimension drawing on
page 9. (The tine orientation marking on the process connection is to identify the
vertical orientation of the fork.)
5.
Shorten sensor cable to a free length of 150 mm (5.9").
6.
Shorten earth cable to a free length of 220 mm (8.7").
7.
Prepare sensor cable as shown above, stripping a maximum of 8 mm from each
wire.
Page EN-18
customer supplied
pipe extension
earth cable (green/yellow)
220 mm
(8.7")
150 mm (5.9")
8 mm max.
18 mm
SITRANS LVS200 – INSTRUCTION MANUAL
terminal crimper
36 mm open end wrench
pipe wrench
fork assembly
7ML19985FT62