Planning/configuring

3.1

Connecting SITRANS RD500 to MultiRanger

1. Connect the RS485 port on the SITRANS RD500 to the MultiRanger RS485 port on TB3 using
wiring and practices in accordance with local electrical codes and requirements.
2. Set up a connection between the PC and SITRANS RD500, referring to the SITRANS RD500
Application instruction manual if necessary. Be sure that you have connected the PC to the
RD500 using the Ethernet crossover cable. (Note: This application guide assumes the
SITRANS RD500 has default network settings.)
6
Configuring SITRANS RD500 for connection to MultiRanger with MODBUS RTU
Application examples, 09/2020, AG111710