5.3 Displayed Messages

Directly above the Metering screens is the Test status display. This portion of the display is
active during the test and allows you to view the type of test being performed and status of the
test step. At the end of a test, the Test status display will either inform you that the test has
passed or give a direct indication of the kind of failure that occurred during the test.

5.3.1 Test Status Messages

Dwell
This message appears on the display while a test is in process, the values update in real time.
Delay
This message appears on the display while a test is in process, the values update in real time.
Pass
This message appears on the display when the test process is complete and the DUT passed
the test.
Abort
This message appears on the display if the test in process is aborted with the RESET button or
remote Reset control.
Short
This message appears on the display if the DUT current is well beyond the metering range of
the test.
Output-Error
This message appears on the display, if the instruments output reading does not match the
setting. This message will only be seen if the Exit key is press at the Output Error screen.
If the instrument has an output problem when the TEST button is pressed, the Output Error
screen will appear .
The RESET button is not active in this situation. Only the Exit soft key will allow you to
return to the Perform Test screen.
OTP-Fail
This message appears on the display if the Power Amplifier has overheated. This is abnormal
condition; please call to the factory for assistance.