Troubleshooting; Error Codes - Thermo Scientific Cimarec HP130914 Operation Manual And Parts List

Table of Contents

Advertisement

Troubleshooting

Error Codes

The following errors should not be addressed by the user. If any of the errors appear, contact Technical
Service at 800-553-0039.
Displayed Message
E01
Call Tech. Service
E02
Call Tech. Service
14
Intended to Detect
Thermocouple
out of range.
Excessive heat-up time.
Cause
Thermocouple not connected.
Thermocouple open circuit.
Thermocouple short circuit.
Failure of element.
Failure of element circuit.
Solution
Replace element
assembly (attached to
thermocouple).
Remove short.
Replace element
assembly (attached to
thermocouple).
Replace Element.
Replace Control Board.

Hide quick links:

Advertisement

Table of Contents
loading