Keysight u1701b Quick Start Manual

Dual display handhel capacitance meter
Hide thumbs Also See for u1701b:

Advertisement

Keysight U1701B
Dual Display Handheld
Capacitance Meter
Quick Start Guide

Advertisement

Table of Contents
loading

Summary of Contents for Keysight u1701b

  • Page 1 Keysight U1701B Dual Display Handheld Capacitance Meter Quick Start Guide...
  • Page 3: Keysight U1701B Dual Display Handheld Capacitance Meter

    Certificate of Calibration If anything is missing or damaged, please contact the nearest Keysight Sales Office. For more detailed information, please refer to the Keysight U1701B Dual Display Handheld Capacitance Meter User’s Guide on Keysight Web site (www.keysight.com/find/handheld-tools). To avoid damage to the device, do not exceed the input WA RN ING limit.
  • Page 4: Capacitance Measurement

    5 Read the measurement on the display. To avoid possible damage to the meter or the equipment CAU TI O N CAU TI O N under test, disconnect circuit power and discharge the capacitor before measuring capacitance. U1701B Quick Start Guide...
  • Page 5: Features And Functions

    Set high/low limits for compare mode Static recording mode Compare mode Tolerance mode Relative mode HOLD Data hold SAVE To store the setting value into the memory RANGE Manual range AUTO Auto range HI/LO High/Low limits Backlight display U1701B Quick Start Guide...
  • Page 6: How To Enter Setup Mode

    • The beeper will beep when a new MAX or MIN value has been recorded. • The static recording captures stable values and updates the memory. It will not record values that are overloaded, OL or below 10 count value. U1701B Quick Start Guide...
  • Page 7 To enable the tolerance mode and to set the display Press value as a standard reference • TOL annunciator will be displayed. • The tolerance will be displayed on the secondary display. • The instrument range will be locked. U1701B Quick Start Guide...
  • Page 8 • The secondary display will indicate C01 to C25 according to the comparison record that has been selected. To save comparison set for next entry Press hold for more than 1 s To exit compare mode Press U1701B Quick Start Guide...
  • Page 9 (H01) setting. To exit the HI/LO limit setting mode Press more than 1 s To toggle backlight ON/OFF Press Backlight turns off automatically after setting period by setup mode. hold for more than 1 s U1701B Quick Start Guide...
  • Page 10: Safety Notices

    Caution, risk of electric shock Caution, risk of danger (refer to the instrument manual for specific Warning or Caution information) For further safety information details, refer to the Keysight U1701B Dual Display Handheld Capacitance Meter User’s Guide.
  • Page 11 This information is subject to change without notice. © Keysight Technologies 2009, 2012, 2014 Edition 3, August 2014 *U1701-90047* U1701-90047 www.keysight.com...