Safety and Installation Instructions
404014050 Accuracy
Accuracy is given as
±
(% measurement
+
% of range)
Range
Frequency
(Hz)
24 Hour
(23 ±1 DC)
90 Days
(23 ±S DC)
1 Year
(23 ±S DC)
Temperature
Coefficient! DC
Outside 18 to 28 DC
100~
3 - 5 Hz
1.1 + 0.06
1.1 + 0.06
1.1 + 0.06
0.2 + 0.006
5 -10 Hz
0.35 + 0.06
0.35 + 0.06
0.35 + 0.06
0.1 + 0.006
10Hz-5kHz
0.15+0.06
0.15+0.06
0.15 + 0.06
0.015 + 0.006
5-10kHz
0.35 + 0.7
0.35 + 0.7
0.35 + 0.7
0.03 + 0.006
1 rnA
3 - 5 Hz
1.0 + 0.04
1.0 + 0.04
1.0 + 0.04
0.1 + 0.006
5 -10 Hz
0.3 + 0.04
0.3 + 0.04
0.3 + 0.04
0.035 + 0.006
10 Hz - 5 kHz
0.1 + 0.04
0.1 + 0.04
0.1 + 0.04
0.015 + 0.006
5 -10 kHz
0.2 + 0.25
0.2 + 0.25
0.2 + 0.25
0.03 + 0.006
10 rnA
3-5 Hz
1.1 + 0.06
1.1 + 0.06
1.1 + 0.06
0.2 + 0.006
5 -10 Hz
0.35 + 0.06
0.35 + 0.06
0.35 + 0.06
0.1 + 0.006
10 Hz - 5 kHz
0.15+0.06
0.15 + 0.06
0.15+ 0.06
0.015 + 0.006
5-10kHz
0.35 + 0.7
0.35 + 0.7
0.35 + 0.7
0.03 + 0.006
100mA
3 - 5 Hz
1.0 + 0.04
1.0 + 0.04
1.0 + 0.04
0.1 + 0.006
5 -10 Hz
0.3 + 0.04
0.3 + 0.04
0.3 + 0.04
0.035 + 0.006
10 Hz - 5 kHz
0.1 + 0.04
0.1 + 0.04
0.1 + 0.04
0.015 + 0.006
5-10 kHz
0.2 + 0.25
0.2 + 0.25
0.2 + 0.25
0.03 + 0.006
400 mA
J31
3 -5 Hz
1.0 + 0.1
1.0 + 0.1
1.0 + 0.1
0.1 + 0.006
5 -10 Hz
0.3 + 0.1
0.3 + 0.1
0.3 + 0.1
0.035 + 0.006
10 Hz -1 kHz
0.1+0.1
0.1 + 0.1
0.1 + 0.1
0.015 + 0.006
1kHz-10kHz
0.2 + 0.7
0.2 + 0.7
0.2 + 0.7
0.03 + 0.006
1 A
121
3- 5 Hz
1.0 + 0.04
1.0 + 0.04
1.0 + 0.04
0.1 +0.006
5 -10 Hz
0.3 + 0.04
0.3 + 0.04
0.3 + 0.04
0.035 + 0.006
10 Hz - 5 kHz
0.1 + 0.04
0.1 + 0.04
0.1 + 0.04
0.015 + 0.006
5 -10 kHz
0.35 + 0.7
0.35 + 0.7
0.35 + 0.7
0.03 + 0.006
3A
111121
3- 5 Hz
1.1 + 0.06
1.1 +0.06
1.1 + 0.06
0.1 + 0.006
5 -10 Hz
0.35 + 0.06
0.35 + 0.06
0.35 + 0.06
0.035 + 0.006
10 Hz - 5 kHz
0.15 + 0.06
0.15+0.06
0.15+0.06
0.015 + 0.006
5 -10 kHz
0.35 + 0.7
0.35 + 0.7
0.35 + 0.7
0.03 + 0.006
10 A
[21
3 - 5 Hz
1.1 +0.06
1.1+0.06
1.1 +0.06
0.1 + 0.006
5 -10 Hz
0.35 + 0.06
0.35 + 0.06
0.35 + 0.06
0.035 + 0.006
10 Hz - 5 kHz
0.15+0.06
0.15 + 0.06
0.15 + 0.06
0.015 + 0.006
5 - 10 kHz
0.35 + 0.7
0.35 + 0.7
0.35 + 0.7
0.03 + 0.006
[1]
Part of
10
A range
[2]
Availab[e only on front panel connectors
[3]
400
mA available in software version
1.0.700.18
or greater only.
400
mA continuously;
550
mA for
2
minutes on,
1
minute off; maximum crest factor
3:1
at
400
mA; specification for current above
329
mA is typical.
"
"
I
28
L

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dmm4050