Compliance Information
This section lists the EMC (electromagnetic compliance), safety, and environmental
standards with which the instrument
complies.
EMC Compliance
EC Declaration of Conformity-EMC
Meets intent of Directive
2004/
108IEC for Electromagnetic
Compatibility.
Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European
Communities:
EN 61326-1 2006, EN 61326-2 2006.
EMC requirements for electrical equipment for
measurement, control, and laboratory use.
1,
2.
3
CISPR 11 :2003. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2:200 l. Electrostatic discharge immunity
IEC 61000-4-3:2002. RF electromagnetic field immunity 4
-
IEC
61000-4-4:2004.
Electrical fast transientlburst immunity
IEC 61000-4-5:200l. Power line surge immunity
IEC 61000-4-6:2003. Conducted RF immunity 5
IEC 61000-4-11:2004. Voltage dips and interruptions immunity 6, 7
EN 61000-3-2:2006.
AC power line harmonic emissions.
EN 61000-3-3:1995.
Voltage changes, fluctuations, and flicker.
European Contact.
Tektronix UK,
Ltd.
Western Peninsula
Western Road
Bracknell, RG 12 lRF
j
United Kingdom
This product is intended for use in nonresidential areas only
,
Use in residential areas may cause electromagnetic
interference,
Emissions which exceed the
levels
required by this standard may occur when this equipment is cormected to a test
object.
To ensure compliance with the EMC standards listed here, high quality shielded interface cables should be used.
The tolerance of the 100uADC, I OmADC ranges while subjected to the test field (3 V im over the frequency range
of80 MHz to I GH z, 1.4 GHz to 2.0 GH
z,
and I Vim from 2.0 GHz to 2.7 GHz, with
80%
amplitude modulation
at I kHz) can increase by
+1-
(0.06%
of range) over the frequency range of 1.7 -
1.9GHz.
(IEC 61000-4-3),
The tolerance of the 100uADC, 10mADC ranges wh iIe subjected to the injected test signal (3 V rms over the
frequency range of 150 kHz to 80 MHz, with
80% amplitude
modulation at I kHz) can increase by
+1-
(0.08%
of
range) at the frequency range of20 - 50MHz. (lEC 61000-4-6).
6
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 00/01250 cycle Voltage-Interruption test
levels (IEC 61000-4-11).
Instrument rebooting may be experienced where the EUT takes approximately 18 seconds to recover from IEC
61000-4-11 transient immunity test.