7.3.2 Access Cover (Optional)

An aluminum access cover (Figure 3) is available for optimum stability. Holes
should be drilled in the access cover to allow insertion of the probes into the well.
The holes must be within the guide ring for the probes to fit into the probe basket.

7.3.3 Probe Basket

A probe basket (Figure 4) is provided as a guide for the probes and to prevent
bumping of the stir bar.

7.3.4 Stir Bar

The stir bar (Figure 5) sits in the bottom of the well for mixing the fluid providing
better accuracy, uniformity, and stability.

7.3.5 Well Extender (Optional)

An optional well extender is available for increasing the depth of the well. This
extender screws onto the well and is equipped with an O-ring. Thread the
extender down until the O-ring makes a good seal. Some fluids may expand up
to 6.35 cm (2.5 inches). Therefore, when using the well extender keep in mind
how much the fluid you are using will expand. Do not overfill the well.
7-4
Find Quality Products Online at:
DO NOT install an access cover without holes (like the optional
cover) onto a bath that is energized. Dangerous pressures may
result from fluids vaporizing.
GlobalTestSupply
www.
Warning:
Figure 4. Probe Basket
Figure 5. Stir Bar