SPECIFICATIONS

Model No.: TD-4242
Dimension & Weight: 96(L) x 50(W) x 23(H) mm, 76.15g
Power Source: two 1.5V size AAA alkaline batteries
Display: LCD
Memory: 450 measurement results with respective date and time
External Output: RS232 PC interface
Auto electrode insertion detection
Auto sample loading detection
Auto reaction time count-down
Auto switch-off after 3 minutes without action
Temperature Warning
Operating Conditions:
50 ° F to 104 ° F (10 ° C to 40 ° C), below 85% R.H. (non-condensing)
Storage / Transportation Condition:
-4 ° F to 140 ° F (-20 ° C to 60 ° C), below 95% R.H.
Measurement Units: mg/dL
Measurement Range: 20 to 600mg/dL (1.1 to 33.3mmol/L)
This device has been tested to meet the electrical and safety
requirements of: IEC/EN 61010-1, IEC/EN 61010-2-101, IEC/EN
61326-1, IEC/EN 61326-2-6