Siemens SOMATOM Emotion 6 Application Manual page 95

Table of Contents

Advertisement

Neck
kV
Effective mAs
Slice collimation
Slice width
Feed/Rotation
Rotation time
Kernel
Increment
Image order
CTDI
Vol
nd
For the 2
reconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
MIPthin: AngioCarotid
Image thickness
Distance between images
Number of images
Contrast medium IV injection
Start delay
Flow rate
Total amount
CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the aortic arch
with triggering threshold of 120 HU, or use manual
triggering.
AngioCarotid
110
70
1.0 mm
3.0 mm
9.0 mm
0.8 sec.
B31s
3.0 mm
cr-ca
6.4 mGy
15 sec.
4 ml/sec.
90 ml
2
nd
reconstr.
1.25 mm
B31s
0.8 mm
coronal
3
3
20
95

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents