Innerear - Siemens SOMATOM Emotion 6 Application Manual

Table of Contents

Advertisement

Head

InnerEar

Indications:
Spiral mode for inner ear High Resolution studies,
e. g. inflammatory changes, tumorous processes
of pyramids, cerebellopontine angle tumors, post-
traumatic changes, etc.
A range of 40 cm will be covered in 15.3 sec.
kV
Effective mAs
Slice collimation
Slice width
Feed/Rotation
Rotation time
Kernel
Increment
Image order
CTDI
Vol
Contrast medium IV injection
Start delay
Flow rate
Total amount
For image reconstruction of soft tissue, use kernel
H30s/H31s.
88
InnerEar
130
150
0.5 mm
0.63 mm
3.0 mm
1.0 sec.
H90s
0.5 mm
ca-cr
42.7 mGy
60 sec.
2 ml/sec.
50 ml