Head
kV
mAs
Slice collimation 2.0 mm
Slice width
Feed/Scan
Rotation time
Kernel
Image order
CTDI
Vol
PosteriorFossaOptimization Filter
PFO: To reduce beam-hardening artifacts in head
images, use for both ranges the Posterior Fossa
Optimization (PFO) filter.
Contrast medium IV injection
Start delay
Flow rate
Total amount
BaseSeq
130
260
4.0 mm
12.5 mm
1.5 sec.
H31s
ca-cr
58.8 mGy
60 sec.
2 ml/sec.
50 – 60 ml
CerebrumSeq
130
300
3.0 mm
6.0 mm
19.0 mm
1.5 sec.
H31s
ca-cr
63.2 mGy