Siemens SOMATOM Emotion 6 Application Manual page 79

Table of Contents

Advertisement

Head
kV
Effective mAs
Slice collimation
Slice width
Feed/Rotation
Rotation time
Kernel
Increment
Image order
CTDI
Vol
For the 2
nd
reconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
MIPthin: AngioHead
Image thickness
Distance between images
Number of images
Contrast medium IV injection
Start delay
Flow rate
Total amount
Use of CARE Bolus with monitoring scans positioned at
the level of the basilar artery or carotid artery. Set the
trigger threshold at 120 HU, or use manual triggering.
AngioHead
110
70
1.0 mm
3.0 mm
6.0 mm
0.8 sec.
H31s
3.0 mm
ca-cr
13.3 mGy
18 sec.
3.5 ml/sec.
75 ml
2
nd
reconstr.
1.25 mm
H20s
0.6 mm
coronal
3
3
35
79