Pelvis
kV
Effective mAs
Slice collimation
Slice width
Feed/Rotation
Rotation time
Kernel
Increment
Image order
CTDI
Vol
nd
For the 2
reconstruction the Autoload into the MPR
Range on the 3D Card is activated. The images will
be automatically loaded into 3D, MPR, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
MPRthick: Hip
Image thickness
Distance between images
Number of images
In case of 3D study only, mAs can be reduced up to
50%, images should be reconstructed with at least 50%
overlapping and kernel B10s.
If only one side is examined, it is advisable to enter the
side in the comment line on the routine card.
Hip
130
95
1.0 mm
5.0 mm
9.0 mm
1.0 sec.
B41s
5.0 mm
cr-ca
12.5 mGy
2
nd
reconstr.
1.25 mm
B70s
0.8 mm
coronal
3
3
20
169