Siemens SOMATOM Emotion 6 Application Manual page 157

Table of Contents

Advertisement

Abdomen
kV
Effective mAs
Slice collimation
Slice width
Feed/Rotation
Rotation time
Kernel
Increment
Image order
CTDI
Vol
st
For the 1
reconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
MIPthin: AngioFast
Image thickness
Distance between images
Number of images
Contrast medium IV injection
Start delay
Flow rate
Total amount
* CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.
Do not administer oral contrast medium, as this
impairs the editing of MIP/SSD/VRT images.
AngioFast
110
90
2.0 mm
3.0 mm
21.0 mm
0.8 sec.
B31s
3.0 mm
cr-ca
7.0 mGy
20 sec.*
3.5 ml/sec.
120 ml
coronal
3
3
35
157