General
How to Create your own
Scan Protocols
User-specific scan protocols can be saved with the
following basic procedure:
• Register a test patient, patient position must be
Head First – Supine
• Select a scan protocol.
• Set the table position to 0 (either at the gantry panel,
or on the topogram routine card).
• Modify the scan protocol, change parameters, add
new ranges etc.
• Do not load the scan protocol.
• Select Edit/Save Scan Protocol in the main menu.
• Select the organ region and the scan protocol name
in the pop-up dialog. You can either use the same
name to modify the existing scan protocol, or enter
a new name, which will create a new protocol name
and will not alter any of the existing protocols already
stored.