For example only
If you are using the normal control solution, according to this particular container of strips,
your meter should show a number between 4.3 and 6.7mmol/l.
H.
14.2-19.1
N.
4.3-6.7
L.
1.9-3.6
Please note that the control values shown in the picture are not control ranges.
NOTE
- It is recommended that glucose control solution is stored at room temperature 20~25
℃ (68~77℉) before testing.
- Check the expiration date before performing a glucose control solution test.
- Do not use beyond expiration date. Once opened, the control solution expires after 3 months.
If your control solution test falls out of range, please follow the next steps before
contacting customer support:
- Check the expiration dates on all the products you are using. Be sure you are using
Finetest™ Lite Control Solution.
- Try another control solution test.
If this test falls out of range, try another control solution test with a new unopened
container of test strips.
If you continue to get control soulution test results that fall outside the range, do not
perform a glucose test. please call Neon Diagnostics Ltd.
• 18 •