Infopia Finetest Lite Instructions For Use Manual page 18

Blood glucose monitoring system for self testing
Table of Contents

Advertisement

For example only
If you are using the normal control solution, according to this particular container of strips,
your meter should show a number between 4.3 and 6.7mmol/l.
H.
14.2-19.1
N.
4.3-6.7
L.
1.9-3.6
Please note that the control values shown in the picture are not control ranges.
NOTE
- It is recommended that glucose control solution is stored at room temperature 20~25
℃ (68~77℉) before testing.
- Check the expiration date before performing a glucose control solution test.
- Do not use beyond expiration date. Once opened, the control solution expires after 3 months.
If your control solution test falls out of range, please follow the next steps before
contacting customer support:
- Check the expiration dates on all the products you are using. Be sure you are using
Finetest™ Lite Control Solution.
- Try another control solution test.
If this test falls out of range, try another control solution test with a new unopened
container of test strips.
If you continue to get control soulution test results that fall outside the range, do not
perform a glucose test. please call Neon Diagnostics Ltd.