4. On the UV Setting screen, each setting of UV is
available. Press Apply key to save the input value
and setup. The display returns to the Tools #1
screen.
●Each setting
・UV Timer:
Set value of period to light UV lamp after closing the outer door.
Settable range: 0 minute~30 minutes, factory setting: 10 minutes.
Note:
・It is recommended to set the UV Timer for 10 minutes. The setting for less than 10 minutes may result in
insufficient disinfection.
・When the UV timer is set for 0, the UV lamp does not light.
・UV Life:
The total time which UV lamp has turned on is displayed as the percentage to 5,000 hours which are
recommendation time to replace. (It is impossible to set).
・UV Timer Ext.:
The more total time which UV lamp has turned on increases, the more UV ray output declines. In order to
cover a decline of the UV ray output, the lighting time of UV lamp is automatically extended with an
increase of total lighting time of UV lamp (The set value of UV Timer is not changed).
Extension rate: 0 %~40 % (It is impossible to set), factory setting: 0 %.
Example) UV Timer: 10 minutes, UV Timer Ext.: 40 % → UV lamp lights for 14 minutes.