UV LAMP PARAMETERS
 If the UV lamp burns out, "Err18: UV Lamp Abnormal" is displayed in the message display field. If this
occurs, replace the UV lamp. When replacing the UV lamp, replace the glow starter at the same time.
Contact our sales representative or agent for information on replacing the UV lamp.
 If the UV lamp burns out ("Err18: UV Lamp Abnormal" is displayed in the message display field), it is not
possible to perform H
O
decontamination. Replace the UV lamp and the glow starter.
2
2

Setting UV lamp ON period

Use the following procedure to change the setting of the UV lamp ON period.
1. Press Menu key to lead the Menu screen.
2. Press Tools #1 key to lead the Tools #1 screen.
3. Press UV Setting key to lead the UV Setting
screen.