STD GAS AUTO CALIBRATION KIT (OPTION)

When a STD gas auto calibration kit MCO-SG is installed, by connecting standard gas cylinder for
calibration, it is possible to calibrate CO
1. Connect a standard gas cylinder to connecting
port for standard gas cylinder on lower right side of
the CO
incubator. Since a Standard gas cylinder is
2
used as a standard of exact density during CO
density calibration, prepare a standard gas cylinder
that is same as the set CO
Note: There is not the problem to remain
connected standard gas cylinder after finished CO
gas density calibration.
2. Press Menu key to lead the Menu screen.
3. Press Tools #1 key to lead the Tools #1 screen.
4. Press STD Gas Setting key to lead the STD Gas
Setting screen.
density manually.
2
2
density.
2
2
Lower right side
Connecting port for standard gas cylinder
79