4. On the Auto-Lock screen, each setting of auto
lock is available. Press Apply key to turn the auto
lock ON and save the set value. The display
returns to the Lock screen.
●Each setting of auto lock
・Auto-Lock:
Auto lock function is that the outer door is locked automatically when the setting time passed after the door
closed. By holding the Auto-lock slide key and sliding it right, the Auto-lock is turned to ON. Settable range:
1 minute~60 minutes, Factory setting: 1 minute.
・User-ID:
Choose the mode of unlocking the outer door between the quick mode or the User-ID mode. By holding
the User-ID slide key and sliding it right, the User-ID mode is turned to ON. Factory setting: OFF (quick
mode).
Note:
・When no User-ID is registered, notice dialog box
is displayed. Press OK key, and then register a
User-ID and its password in reference to page 68~
70.
・In the User-ID mode, User-ID which is inputted to
unlock the outer door is saved as the open/close
state of outer door log data (refer to page 45~46).
・ When changing the User-ID mode to OFF,
registered User-IDs are not deleted.
・When deleting all registered User-IDs, the User-ID
mode is turned to OFF (refer to page 70).
5. Press Top key to return to the Top screen.
71