HP 83522A Service Notes page 93

Rf plug-.in including options 002 and 004
Table of Contents

Advertisement

Adjustments
Model 83522A
ADJUSTMENTS
5-27. MARKER AND SAMPLER ADJUSTMENTS (Cont'd)
2. Set 8350A
START/STOP
sweep for 10 MHz to 2.4 GHz. Select 83522A
AMPTD
.
Connect counter- with 1:l capacitive probe to A8TP1. Adjust A8C4
for frequency counter indication of 50 MHz
f
250 Hz. If A8C4 does not
have the range required to adjust the 50 MHz crystal oscillator, select a new value for
A8C3. (An increase in capacitance will decrease frequency.)
1 MHz
10 MHz
50 MHz
1M
10M
50M BIAS
A8R53
A8R54
A8R55
A8TP3
A8R29
A8R30
A8R31
50 MHz OSC.
A8TP1
SELECTED
VALUE
A8C3"
50 MHz OSC.
A8C4
23
YY
Figure 5-33. Marker Adjustments on A 8
3. Select 100 ms Sweep Time. Connect oscilloscope with 1:l probe to A8TP3 (Figure 5-33).
Set 8350A power to +13 dBm and select
1 M H e
Markers. Adjust A8R29 (1M) for the
flattest envelope hei ht. (See Figure 5-34.) Select
tKl r '
Markers. Adjust A8R30 (IOM)
for the flattest envefope height (Figure 5-33). Select
51)
M E 3 2
markers. Adjust A8R31
(50M) for the flattest envelope height. (Optimum setting for these adjustments will be
ones that provide the most uniform birdie height acrossthe band with the adjustments
nearest the center of their range.) Especially note birdie height at the high-frequency
end and set the adjustment just before the marker amplitude drops off.
4. Connect oscilloscope to A8TP4 (Figure 5-33). Set RF
to 0 dBm. Adjust IF
gain potentiometers A8R53 (1 MHz), A8R54 (10 MHz),
5 (50 MHz) for each
marker frequency to an average envelope height of 1.0 V p-p.
5. Adjust oscilloscope Channel B vernier for a horizo
divisions. Set 8350A
Cp
=
1
&+p
=
fO
MWr
Select
birdie envelope on the screen with plug-in front panel
Figure 5-35.) Then select 10 MHz Markers. Change
4LS:
to
Display is now calibrated for 100 kHz1Division.