HP 83522A Service Notes page 240

Rf plug-.in including options 002 and 004
Table of Contents

Advertisement

Service
Model 83522A
Flatness/Oscillations (Power Drop-outs)
Monitor the RF output with the HP 8755C as shown in Fig. 8-26.
If the power level is constant across the sweep within approximately 5 dB,
then the plug-in may only require ALC flatness adjustments. Refer to
Section V, Adjustments, in this manual, for the Internal Leveled Flatness
adjustment procedure.
If the measured power level lies between 4-13 and -2 dBm, but can't be
controlled via the front panel, refer to the Digital Control section under
Troubleshooting Diagnosis.
If the trace appears chopped or broken the loop may be oscillating. Refer to
Section V, Adjustments, in this manual, and perform the ALC Gain
adjustment procedure.
Full Unleveled Power
Set the 83522A to sweep the full range.
Attempt to level the power externally using the HP 432A Power Meter as
shown in Figure 8-27. Select
MTR
leveling, and enter a slow (at least 30
seconds) sweep time. If the RF power is now leveled then the failure is most
likely in the detector or the Detector Selection Switch, A4U6. Refer to the
following paragraph. If this does not prove to be the case, the problem may
be in the two analog switches U4B and U6A. It may be necessary to perform
the ALC adjustments in Section V of this manual.
SWEEP
R
F
OSCILLATOR
PLUG-IN
/
\
POWER
MOUNT
u
Figure 8-2 7. Power Meter Leveling Setup
Check the Detector Selection Switch by entering a CW frequency at the
desired frequency or in the leveling mode in question and trace the detector
voltage through U6B. If the input to be selected doesn't match the output,
check the MUXAO and MUXAl lines (see Table 8-10). Also check U12 and
U13 as described under Digital Control.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents