HP 83522A Service Notes page 71

Rf plug-.in including options 002 and 004
Table of Contents

Advertisement

Adjustments
Model 83522A
ADJUSTMENTS
5-1 8. DELAY COMPENSATION (Confd)
15. On the 83522A, ensure that the power is set to +13 dBm and press
16. Set the oscilloscope controls as follows:
Display Mode
....................................................
A vs.B
Display.
......................................................
MAG X10
Ch. A Vertical Sensitivity.
........................................
2V/DIV
Ch. B Vertical Sensitivity..
.......................................
lV/DIV
Adjust the HOIZONTAL POSITION control to set the start of sweep exactly on the
leftmost graticule.
NOTE
Although the HP 1740A is the specified oscilloscope, the use of an
oscilloscope with a variable persistance screen may be advantageous in
order to more clearly see the 5 0 MHz markers when the sweep speed is
decreased to 10 milliseconds.
17. On the 83504 press
.
Rotate the VERNIER to place the third crystal
marker (150 MHz marker) exactly 0.4 divisions to the left of the leading edge of the
8350A amplitude marker as shown in Figure 5-15. It may be necessary to fine adjust the
oscilloscope horizontal position control and the Channel A vertical position control to
move the leading edge of the 8350A amplitude marker and the desired crystal marker to
a convenient graticule which may be used as a point of reference.
0 Vdc-
-
RETRACE
NOTE
THE MARKER SPACING MAY
VARY SLIGHTLY DUE TO DI-
FERENCES IN OSCILLOSCOPE
HORIZONTAL GAIN WHEN I N
X I 0 MAG MODE.
-50
MHz CRYSTAL MARKERS
8350A MARKER M I
=SET
FOR 0.4 DIVISION SPACING
2
Figure 5-1 5. Delay Compensation Adjustment Wave form