Power Sweep - HP 83522A Service Notes

Rf plug-.in including options 002 and 004
Table of Contents

Advertisement

Model 83522A
ADJUSTMENTS
Adjustments
5-25. POWER SWEEP
REFERENCE:
Performance Test: 8350A Paragraph 4- 14.
Service Sheet: A5
DESCRIPTION:
A 10 dB/sweep power sweep mode is selected and the resultant is displayed on the 8755C
Swept Amplitude Analyzer. Output of the power sweep circuit is adjusted for the correct
sweep.
EQUIPMENT:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sweep Oscillator
HP 8350A
....................................
Swept Amplitude Analyzer.
HP 8755ClHP 182T
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Detector.
HP 11664A
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10 dB Attenuator.
HP 8491A Option 010
PROCEDURE:
NOTE
ALC gain adju-atments (paragraph 5-24) must be checked before power
sweep adjustment are made.
NOTE
This procedure assumes that A3S1 is set to the factory-set position (Table
5-6), and at the 8350A Sweep Oscillator, 27.8 kHz square wave modulation
is selected.
1. Connect equipment as shown in Figure 5-27. Select 8350A