Keithley 3720 User Manual page 155

Switch/multimeter series 3700
Hide thumbs Also See for 3720:
Table of Contents

Advertisement

Series 3700 System Switch/Multimeter User's Manual
Example 2
This example is for a partially loaded Model 3706 with Model 3723 cards (all 1 pole mode).
Slot #
SLOT 1
SLOT 2
SLOT 3
SLOT 4
SLOT 5
SLOT 6
This produces the following power consumption:
Slot 1 Power Consumed =
Slot 2 Power Consumed =
Slot 3 Power Consumed =
Slot 4 Power Consumed =
Slot 5 Power Consumed =
Slot 6 Power Consumed =
Totals for each bank can then be calculated as follows:
Bank #1 Power Consumed =
Bank #2 Power Consumed =
Result: Since each bank has not exceeded maximum power, this is OK.
Document Number: 3700S-900-01 Rev. A / August 2007
Channel relays
Card
closed
3723
107
3723
107
EMPTY
0
3723
107
3723
107
EMPTY
0
700
700
0
700
700
0
Slot 1
6150
Slot 4
6150
Section 9: Series 3700 Module Schematics and Connections
Backplane relays
closed
1
1
0
1
1
0
+
107 x 50
+
1 x 100
+
107 x 50
+
1 x 100
+
0
+
0
+
107 x 50
+
1 x 100
+
107 x 50
+
1 x 100
+
0
+
0
Slot 2
Slot 3
+
6150
+ 0
Slot 5
Slot 6
+
6150
+ 0
=
6150
=
6150
=
0
=
6150
=
6150
=
0
Total
=
12300
Total
=
12300
9-5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

37213722372337303740

Table of Contents