Enabling Math; Additional Math Operations - Keithley 6517B User Manual

Electrometer
Hide thumbs Also See for 6517B:
Table of Contents

Advertisement

Section 4: Measurement Options

Enabling math

The selected math calculation is enabled by simply pressing the MATH key. When enabled, the
MATH annunciator turns on and the calculation type (NONE, POLY, %, or % DEV) is displayed.
Also, one of the following messages is briefly displayed:
Math Enabled
Display = NONE (reading)
Math Enabled
Display = POLY (reading)
Math Enabled
Display = % (reading)
Math Enabled
Display = % DEV (reading)
Math Enabled
Display = DEV (reading)
Math Enabled
Display = RATIO (reading)
Math Enabled
Display = LOG10 (reading)
Math is disabled by pressing MATH a second time.

Additional math operations

In addition to the math performed on single readings described above, the Model 6517B has these
math operations:
Math performed on buffered readings (maximum and minimum values, average, and
standard deviation).
Math performed on single readings as part of a pass/fail limits test.
Complete information on MATH is provided in Section 9 of the Reference Manual.
Figure 4-7
Combining math calculations
4-18
Return to