Error Messages - Keithley 486 Quick Reference Manual

Picoammeter/source
Hide thumbs Also See for 486:
Table of Contents

Advertisement

ERROR MESSAGES
Table 1. Error Messages
Message
Description
NO DEFLTS
Checksum error in default condi-
tion during power-up.
UNCAL
Checksum error in calibration
during power-up
ROM ERROR
Failed ROM self-test.
RAM ERROR
Failed RAM self-test.
TRG OVERUN
Trigger overrun; instrument trig-
gered while still processing a pre-
vious trigger.
CAL VALUE
Calibration value conflicts with
the currently selected range.
CAL ERROR
Calibration value not within allow-
able limits.
CAL LOCK
Sent calibration command with
calibration switch (CAL LOCK)
disabled.
CAL ZCHK
Cannot calibrate picoammeter
with zero check enabled.
INTERLOCK
Tried to put V-Source of Model
487 in operate while an enabled
safety interlock switch is open.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

487

Table of Contents