When zero correction is performed, only the present
range is zero corrected. Before making measurements, it
is recommended that each current range be zero cor-
rected.
To perform zero correction:
1. Enable zero check.
2. To perform zero correction, press SHIFT and then
CORRECT in that order.
Filters
SHIFT
cl
Filtering is used to stabilize noisy measurements. The
Model 486/487 has two available filters; a digital filter and
an analog filter. The digital filter bases the reading on the
weighted average of a number of measurement conver-
sions. The analog filter is a simple RC filter whose time
constant varies with the selected range.
When the filter is enabled, the selected filter(s) (digital fil-
ter, analog filter or both) will be used.