Specifications - Oakton pH 5+ Instruction Manual

Portable meters
Hide thumbs Also See for pH 5+:
Table of Contents

Advertisement

Instruction Manual

6. SPECIFICATIONS

Ion Range
Resolution
Accuracy
Calibration Points
pH Range
Resolution
Accuracy
Slope Range
Calibration Points
Buffer Groups
Temperature Range
Resolution
Accuracy
Compensation
Millivolt Range
Millivolt Range
Resolution
Accuracy
Auto-Buffer Recognition
Hold Function
Auto Shut Off
Low Battery Indication
Operating Temperature
Power Requirements
Battery Life
Meter Dim./Weight
Model
0.01 to 0.99; 1.0 to 199.9; 200 to 1999
0.01 / 0.1 / 1
± 1 % full scale
2 or 3 consecutive points;
(0.1, 1.0, 10.0, 100.0, 1000)
0.00 to 14.00 pH
0.01 pH
± 0.01 pH
80 to 120%
2 to 5 points
1.68, 4.01, 7.00, 10.01, 12.45 (USA)
1.68, 4.01, 6.86, 9.18, 12.45 (NIST)
4.10, 6.97 (Pb)
0.0 to 100.0
o
0.1
C
o
± 0.5
C
Automatic / Manual (0 to 100
-1000 to +1000 mV
-500 to 500 mV
± 1 mV;
± 0.1 mV between –200 to 200 mV
± 2 mV;
± 0.2 mV between –200 to 200 mV
Features
(4) x 1.5V AAA Alkaline Batteries (included)
15.7 x 8.5 x 4.2 cm / 255 g
17
o
C
o
C)
Yes
"HO"
After 17 minutes
"LO"
o
0 to 50
C
500 hours
pH 5+, pH 6+, Ion 6+
pH 5+ pH 6+ Ion 6+

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ph 6+Ion 6+

Table of Contents