Without Temperature Probe (Manual Temperature Compensation)

If a temperature probe is not connected, the meter compensates for pH response based
on the factory default value at 25.0 °C. This default temperature can be manually
adjusted using the procedure below.
For nearly all applications however, Automatic Temperature Compensation (ATC) is
recommended for best accuracy.
1.
Turn on the meter and select °C mode by pressing
the temperature probe.
2.
Press
briefly. The factory default temperature (25.0 °C value will blink.
3.
Use
(pH 5+ and pH 6+) or
desired value. The maximum adjustment is ± 0 to 100 °C.
4.
Press
automatically save the calibration, cease blinking, and begin °C measurement with
the new default temperature.
to begin temperature calibration mode. "" (calibrate) will display
to confirm the value. "" (confirm) will display briefly. The meter will
(Ion 6+) to adjust the reading to the