Operating Instructions OAKTON ® pH 10 series pH/mV/°C Meter MEAS READY pH • mV • C Meter pH 10 series HOLD MODE MEAS Printed in Singapore 8/00 rev 1...
Table of Contents 1. Introduction ................Page 4 Getting started • Maintenance, troubleshooting 2. Display and keypad functions..........Page 5 3. Preparation ................Page 6 • Inserting the batteries • Connecting the electrode and temperature probe • Attaching the electrode holder to the meter •...
1. Introduction 2. Display and Keypad functions Thank you for selecting an OAKTON ® meter. The OAKTON ® pH 10 series The water-resistant membrane meters are economical, microprocessor-based meters that deliver high ±0.01 pH keypad makes the instrument easy MEAS accuracy.
4. pH Calibration 6. Press CON to confirm calibration. A confirming indicator flashes for one second and disappears. See Figure I. The meter is This instrument is capable of up to 3-point calibration to ensure accuracy READY now calibrated at the buffer indicated in the across the entire range of the meter.
Measurement with automatic temperature compensation (ATC) 2. Turn the meter on. Make sure the meter is in the Measurement Mode Automatic temperature compensation requires an OAKTON “All-in-One” (MEAS indicator shows on the top center of the LCD). If the meter is not in electrode or a separate temperature probe.
6. When the reading stabilizes, the READY indicator will display. See Figure O. NOTE: The pH 10 series meter does not have a relative mV mode. We have other meters that can take relative mV readings—contact your OAKTON distributor for information. Page 12 Page 13...
8. Temperature calibration 2-point temperature calibration In most cases 1-point temperature calibration will give you accurate 1-point temperature calibration temperature measurement and compensation. If you notice inaccurate measurements at high or low temperatures after 1-point calibration, perform the 1. Connect ATC probe (or temperature connector of the “All-in-One” following procedure.
C. Temperature Low Point Calibration 9. Electrode care and maintenance The primary display will show "0.0" and Because your pH electrode is susceptible to dirt and contamination, clean it the secondary display will show "r2.0". every one to three months depending on the extent and condition of use. See Figure U.
* not improve, then your electrode is no longer functioning. Replace with a new Turn meter off and Err 3 in electrode. Call your OAKTON distributor for more information. ADC error Hardware failure then on. Return if primary display necessary.
Compensation 0 to 100°C WD-00654-08 pH 10.01 calibration buffer; 1 pint, 1.3 lbs/0.59 kg pH slope range: 80% to 120% Operating Temperature: 0 to 50°C Contact your OAKTON ® distributor for ordering information and for a (32° to 122°F) ATC function: 0 to 100°C complete selection of pH and ORP electrodes, solutions, holders, and Power: 4 x 1.5 V AAA batteries (50 hrs);...
For your reference, please note in this space below the name of your shipment reference number and your purchase order number. OAKTON warrants this product to be free from significant deviations in material and workmanship for a period of one year from date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the one year period, please return—freight prepaid—and correction will be...