Repairs
Mounting
1. Mount the seat angle mechanism using the four screws. See the illustration. It can be mounted in
three different positions, depending on which chassis it is being mounted on and the depth of the
seat. See the table below. Use a torque wrench to tighten the screws. Tightening torque 24 Nm.
Seat depth
370 mm./15"
395 mm./16"
420 mm./17"
445 mm./18"
470 mm./19"
495 mm./20"
520 mm./21"
545 mm./22"
570 mm./23"
The mounting position for the seat angle mechanism depends on the chassis model and the seat depth
setting. The table shows mounting positions for standard model wheelchairs; accessories such as
height-adjustable footplates or a respirator shelf may affect the mounting position. Contact Permobil in
case of doubt.
The seat angle mechanism is held in place by four screws