Service
Function of the ) Key
(> 2 sec)
Purpose
Normally, when you press and hold
the ) key for longer than 2 seconds,
the calibration/adjustment function
is triggered. When the service mode
is active, you can assign a different
function to this key:
Diagram of the Menu for Assigning the Function of the ) Key (> 2 sec)
1. 9.
Calibration and Adjustment
1. 9. 1
External calibration/adjustment with default weights (service mode not required)
1. 9. 3
External calibration/adjustment with user-defined weights entered under 1.18 (service mode not required)
1. 9. 5
Internal linearization (WP2 only)
1. 9. 6
External linearization with default weights
1. 9. 7
External linearization with user-defined weights (enter under 1.18)
1. 9. 8
Set preload
1. 9. 9
Clear preload
1. 9. 10
Key blocked
84
– External linearization with default
weights (menu item 1.9.6)
– External linearization with the
linearization weights specified under
menu item 1.18 (menu item 1.9.7)
– Internal linearization (WP2 only)
(menu item 1.9.5)
– Set the preload (menu item 1.9.8)
– Clear the preload (menu item 1.9.9)
!
Following linearization or after
a preload is set or cleared, the normal
function of the ) key (> 2 sec)
must be reset in the Setup menu (e.g.,
external calibration/adjustment with
default weights)