instructional notes

1. If in setup mode when the test strip is inserted, the meter saves all the values entered up to that point
and immediately switches to test mode. Upon exiting test mode the meter screen goes blank and does
not return to setup mode.
2. If in Data Review mode when the test strip is inserted, the meter immediately switches to test mode.
Upon exiting test mode the meter screen goes blank and does not return to Data review mode.
3. Battery low icon is displayed in every mode except setup.
4. Once battery level goes below the threshold that triggers the "low battery" warning, it continues to give the
warning until the meter becomes unusable due to low battery.
5. The meter responds to the pressing and the holding of keys:
left/right
buttons
- The Left/Right button moves forward/backward through a series of stored test result screens or
increments of value.
- Hold down the Left/Right button to speed up screen change process.
MOde
button
- When the MODE button is pressed for less than 1.5 seconds to advance to the next function, the
meter advances to the next screen immediately when the button is pressed.
- While the meter is in sleep mode (OFF), pressing the MODE button for less than 1.5 seconds wakes
up the meter and enters data review mode.
- While the meter is in sleep mode (OFF), pressing the MODE button for longer than 3 seconds wakes
up the meter and enters setup mode.

appendix
GlucoMen
LX - 38973 - 09/07
®
- While the meter is awake (ON), pressing the MODE button for longer than 1.5 seconds manually turns
off the meter (sleep mode).
6. With no activity, time-out occurs after the following times:
- 1 minute for all screens.
- 2 minutes during test mode.
- 3 minutes when download connector inserted.
GlucoMen
appendix
LX - 38973 - 09/07
®